FDA Adverse Event Malfunction Summary report: N

ROUND FILTER POLYPROVPYLENE

MDR report key: 3196807 · Received June 25, 2013

Report

Report Number
2916714-2013-00036
Event Type
Malfunction
Date Received
June 25, 2013
Date of Event
March 29, 2013
Report Date
May 7, 2013
Manufacturer
SPS MEDICAL SUPPLY CORP
Product Code
FRG
PMA / PMN Number
K040865
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ADD'L 510 (K): K093493, K093649 AND K112671. EVALUATION: THE RETURNED FILTER FROM THE TOP OF THE CONTAINER SHOWED THAT IT WAS QUITE DISTORTED AND HAD SEEN A TEMPERATURE HIGHER THAN THE 270 DEGREES FAHRENHEIT THAT THE STERILIZER WAS SET AT. WITH THE FACT THAT IT WAS THE TOP FILTERS AND NO OTHERS; THE MOST PROBABLE CAUSE WAS STERILIZER OVERLOAD. THE TOP OF THIS CONTAINER WITH THIS FILTER WAS VERY CLOSE OR TOUCHING THE TOP OF THE CHAMBER WHERE THE TEMPERATURE IS MUCH HOTTER. DETERMINED TO BE USER RELATED AND NOT PRODUCT RELATED.

Description of Event or Problem · 1

TOP FILTER WAS APPEARING TO BE MELTED WHEN THE SET WAS OPENED IN THE OPERATING ROOM. BOTTOM FILTER APPEARED FINE AND NO OTHER SETS HAVE HAD THIS MELTING PROBLEM. NO PT INJURY. SURGERY WAS DELAYED 30 MINUTES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
289154 ROUND FILTER POLYPROVPYLENE FRG SPS MEDICAL SUPPLY CORP MD344 1302

Patients

Seq Age Sex Outcome Treatment
1 Other