10MM/130 DEG TI CANN TROCH FIXATION NAIL 170MM-STERILE
Report
- Report Number
- 1719045-2013-01735
- Event Type
- Injury
- Date Received
- June 28, 2013
- Date of Event
- June 1, 2013
- Report Date
- June 1, 2013
- Manufacturer
- SYNTHES MONUMENT
- Product Code
- HSB
- PMA / PMN Number
- K011857
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MO, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
DEVICE IS USED FOR TREATMENT, NOT DIAGNOSIS. THE RETURNED AIMING ARM (PART 357.366, LOT 3360761) WAS MANUFACTURED IN MARCH 2010 AND IS OVER 3 YEARS OLD. THE RETURNED BLADE GUIDE SLEEVE (PART 357.369, LOT 5060068) WAS MANUFACTURED IN NOVEMBER 2005 AND IS OVER 7 YEARS OLD. THE RETURNED BUTTRESS/COMPRESSION NUT (PART 357.371, LOT 4546699) WAS MANUFACTURED IN MAY 2003 AND IS OVER 10 YEARS OLD. THE RETURNED HELICAL BLADE COUPLING SCREW (PART 357.377, LOT 4728595) WAS MANUFACTURED IN JULY 2004 AND IS APPROXIMATELY 9 YEARS OLD. THE RETURNED HELICAL BLADE INSERTER (PART 357.372, LOT 4815961) WAS MANUFACTURED IN JULY 2004 AND IS APPROXIMATELY 9 YEARS OLD. THE RETURNED HELICAL BLADE (PART 456.304, LOT 7050256) WAS MANUFACTURED IN OCTOBER 2012 AND IS LESS THAN 1 YEAR OLD. THE RETURNED SHORT TROCHANTERIC FIXATION NAIL (PART 456.315, LOT 7291560) WAS MANUFACTURED IN MARCH 2013 AND IS LESS THAN A YEAR OLD. THE RETURNED INSERTION HANDLE (PART 357.411, LOT 6286402) WAS MANUFACTURED IN APRIL 2010 AND IS OVER 3 YEARS OLD. DURING THIS EVALUATION, THE RETURNED DEVICES FROM THIS COMPLAINT WERE ASSEMBLED WITH KNOWN GOOD MATING INSTRUMENTS TO COMPLETE THE INSERTION CONSTRUCT IN ORDER TO EVALUATE THE ALIGNMENT OF THE HELICAL BLADE WITH RESPECT TO THE TROCHANTERIC FIXATION NAIL. THE MATING KNOWN GOOD PARTS INCLUDED A CANNULATED CONNECTING SCREW AND BALL HEX SCREWDRIVER. THE CONSTRUCT ASSEMBLED AND THE RETURNED HELICAL BLADE ALIGNED AND ENTERED THE PROXIMAL LOCKING HOLE IN THE RETURNED TROCHANTERIC FIXATION NAIL ON EACH ATTEMPT AS INTENDED. THE COMPLAINT CONDITION COULD NOT BE REPLICATED DURING THIS EVALUATION. THE TROCHANTERIC FIXATION NAIL RISK ANALYSIS ADEQUATELY ADDRESSES THIS COMPLAINT CONDITION AS LESS SEVERE. SPECIFICALLY, THE RISK ASSOCIATED WITH THE 3RD HAZARD LISTED UNDER INSTRUMENTATION ON ASSEMBLY (INSTRUMENTATION DOES NOT TARGET THE IMPLANTS (BOTH PROXIMAL AND DISTAL LOCKING) DUE TO DIMENSIONAL MISMATCH OF HANDLE, AIMING ARM, COMPONENTS OR DAMAGED PRIOR TO USE IS ASSESSED AS LESS SEVERE WITH A MODERATE SEVERITY OF HARM 3 AND AN IMPROBABLE PROBABILITY OF OCCURRENCE 1 WITH POSSIBLE HARM LISTED AS CANNOT USE SYSTEM FOR SURGERY. THE COMPLAINT CONDITION COULD NOT BE REPLICATED DURING THIS EVALUATION. THE CONSTRUCT ASSEMBLED AND THE RETURNED HELICAL BLADE ALIGNED AND ENTERED THE PROXIMAL LOCKING HOLE IN THE RETURNED TROCHANTERIC FIXATION NAIL ON EACH ATTEMPT AS INTENDED. THE DESIGN IS ADEQUATE FOR ITS INTENDED USE AND DID NOT CONTRIBUTE TO THIS COMPLAINT CONDITION. PLACEHOLDER.
DEVICE IS USED FOR TREATMENT, NOT DIAGNOSIS ADDITIONAL CODE: HWC. INVESTIGATION IS ON-GOING. SUBJECT DEVICE HAS BEEN RECEIVED AND IS CURRENTLY IN THE EVALUATION PROCESS. NO CONCLUSION CAN BE DRAWN. A REVIEW OF SYNTHES DEVICE HISTORY RECORDS FOR MANUFACTURING REVEALED NO COMPLAINT RELATED ISSUES.
DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. ADDITIONAL NARRATIVE: A FULL DIMENSIONAL INSPECTION SHOWED ONLY THE ABOVE NOTED ISSUES. THESE ISSUES CAUSED BY SURGERY WERE MINOR AND WOULD NOT CAUSE A FAILURE OF THE NAIL. IN ADDITION, A BENCH FIT CHECK WITH THE RETURNED INSTRUMENTS REVEALED NO ALIGNMENT ISSUES. IT WAS NOTED THAT PART NUMBER 357.397 CONNECTING SCREW WAS NOT RETURNED WITH THE INSTRUMENT ASSEMBLY. A STANDARD SCREW WAS USED FOR THE BENCH FIT. VERIFIED MATERIAL TIALB WITH NITON 07, PASSED.
DURING A PROCEDURE USING THE SHORT TROCHANTERIC FIXATION NAIL AND HELICAL BLADE; THE NAIL WAS IMPLANTED BUT THE BLADE WOULD NOT GO THROUGH THE NAIL. IT WAS IMPACTING THE LATERAL EDGE OF THE HOLE IN THE NAIL. THE SURGEON REMOVED THE NAIL AND BLADE, CHECKED AND RETIGHTENED ALL INSTRUMENTATION, RE-CHECKED ANGULATION OF THE AIMING ARM AND INSERTION HANDLE. THE LOCK IN MECHANISM WAS NOT DEPLOYED. SURGICAL DELAY TIME OF 40 MINUTES WAS REPORTED. THIS REPORT IS 1 OF 8 FOR COMPLAINT (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 296295 | 10MM/130 DEG TI CANN TROCH FIXATION NAIL 170MM-STERILE | HSB | SYNTHES MONUMENT | 7291560 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 75 YR |