FDA Adverse Event Injury Summary report: N

CHARGING SYSTEM

MDR report key: 3193626 · Received June 25, 2013

Report

Report Number
1627487-2013-15847
Event Type
Injury
Date Received
June 25, 2013
Date of Event
June 3, 2013
Report Date
June 3, 2013
Manufacturer
ST. JUDE MEDICAL - NEUROMODULATION
Product Code
LGW
PMA / PMN Number
P010032
Removal / Correction Number
1627487-07262012-001-C
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THIS CHARGER MODEL WAS ASSOCIATED WITH A FIELD CORRECTION. MANUFACTURER'S EVALUATION: CORRECTIVE AND PREVENTIVE ACTION (CAPA) INVESTIGATION WAS PERFORMED. EVALUATION: RESULTS: POCKET HEATING WAS CONFIRMED. THE INVESTIGATION FOR CAPA (B)(4) ASSOCIATED WITH HEATING WHILE CHARGING (POCKET HEATING) CONCLUDED THAT THE CHARGER WAS CAPABLE OF TRANSFERRING ENERGY TO THE IPG AT A RATE THAT WOULD CAUSE HEATING OF THE IPG AND/OR CHARGING WAND OF SUFFICIENT ELEVATED TEMPERATURE TO CAUSE PAIN AND BURNS. THE HEATING WHILE CHARGING WAS DETERMINED TO BE EXACERBATED BY OFF-AXIS CHARGING OF SHALLOW IMPLANTED IPGS AND THAT ALL CHARGERS WERE CAPABLE OF ELEVATED HEATING. SJM HAS LIMITED INFORMATION RELATED TO THE PATIENT'S MEDICAL HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PATIENT'S HISTORY TO THE EVENT REPORTED. SJM DEFERS TO THE PATIENT'S PHYSICIAN REGARDING MEDICAL HISTORY.

Description of Event or Problem · 1

DEVICE 2 OF 2. REFERENCE MFR REPORT: 1627487-2013-15846.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
287528 CHARGING SYSTEM SCS CHARGING SYSTEM LGW ST. JUDE MEDICAL - NEUROMODULATION 3721 3083186

Patients

Seq Age Sex Outcome Treatment
1 58 YR Other SCS EXTENSION: MODEL 3386| IMPLANT:| IMPLANT:| IMPLANT:| SCS LEAD: MODEL 3186| IMPLANT:| SCS LEAD: MODEL 3186| SCS EXTENSION: MODEL 3346