CHARGING SYSTEM
Report
- Report Number
- 1627487-2013-13974
- Event Type
- Injury
- Date Received
- June 25, 2013
- Date of Event
- May 31, 2013
- Report Date
- May 31, 2013
- Manufacturer
- ST. JUDE MEDICAL - NEUROMODULATION
- Product Code
- LGW
- PMA / PMN Number
- P010032
- Removal / Correction Number
- 1627487-07262012-001-C
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WV, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
THIS CHARGER MODEL WAS ASSOCIATED WITH A FIELD CORRECTION. CORRECTIVE AND PREVENTIVE ACTION (CAPA) INVESTIGATION WAS PERFORMED. EVALUATION RESULTS: POCKET HEATING WAS CONFIRMED. THE INVESTIGATION FOR CAPA (B)(4) ASSOCIATED WITH HEATING WHILE CHARGING (POCKET HEATING) CONCLUDED THAT THE CHARGER WAS CAPABLE OF TRANSFERRING ENERGY TO THE IPG AT A RATE THAT WOULD CAUSE HEATING OF THE IPG AND/OR CHARGING WAND OF SUFFICIENT ELEVATED TEMPERATURE TO CAUSE PAIN AND BURNS. THE HEATING WHILE CHARGING WAS DETERMINED TO BE EXACERBATED BY OFF-AXIS CHARGING OF SHALLOW IMPLANTED IPGS AND THAT ALL CHARGERS WERE CAPABLE OF ELEVATED HEATING. SJM HAS LIMITED INFORMATION RELATED TO THE PATIENT'S MEDICAL HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PATIENT'S HISTORY TO THE EVENT REPORTED. SJM DEFERS TO THE PATIENT'S PHYSICIAN REGARDING MEDICAL HISTORY.
DEVICE 2 OF 2. REFERENCE MFR. REPORT: 1627487-2013-13973.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 289197 | CHARGING SYSTEM | SCS CHARGING SYSTEM | LGW | ST. JUDE MEDICAL - NEUROMODULATION | 3721 | 2833641 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 60 YR | Other | IMPLANT:| SCS LEAD: MODEL 3186 (2)| IMPLANT:| SCS ANCHOR: MODEL 1194 (2) |