SCRDRIVERSHAFT F/LOCKCAP F/URS
Report
- Report Number
- 8030965-2013-04056
- Event Type
- Malfunction
- Date Received
- June 27, 2013
- Date of Event
- November 30, 2011
- Report Date
- January 18, 2012
- Manufacturer
- SYNTHES GMBH
- Product Code
- HXX
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KU
- Reporter Occupation
- OTHER
Narratives
SYNTHES IS SUBMITTING THIS REPORT AS A RESULT OF REMEDIATION ACTIVITIES RELATED TO FDA WARNING LETTER DATED FEBRUARY 2012. DEVICE(S) LISTED IN THIS REPORT IS (ARE) USED FOR TREATMENT, NOT DIAGNOSIS. ANY ADDITIONAL INFORMATION RECEIVED REGARDING THIS EVENT AFTER FILING THIS REPORT SHALL BE FILED ON A SUPPLEMENTAL MDR. DEVICE IS NOT DISTRIBUTED IN THE UNITED STATES, BUT IS SIMILAR TO DEVICE MARKETED IN THE USA. THE MANUFACTURING DOCUMENTS WERE REVIEWED AND NO DISCREPANCIES TO THE SPECIFICATIONS WERE FOUND. THE INVESTIGATION OF THE COMPLAIN INSTRUMENTS HAS SHOWS THAT THE TIP OF THE DRIVER IS INDEED BROKEN OFF. THE STRUCTURE OF THE BROKEN SURFACE INDICATES THAT TOO MUCH APPLIED MECHANICAL FORCE HAS LEAD TO THIS DAMAGE. BASED ON THESE FINDINGS WE CONCLUDE THAT THE CAUSE OF FAILURE IS NOT DUE TO ANY MANUFACTURING NON-CONFORMANCES. WE SUPPOSE THAT A MECHANICAL OVERLOAD BY APPLYING TOO MUCH LATERAL STRESS CAUSED THIS BREAKAGE. NO PRODUCT FAULT COULD BE DETECTED.
IT WAS REPORTED THAT THE TIP OF THE SCREWDRIVER SHAFT IS BROKEN OFF. DURING THE FINAL TIGHTENING OF AN URS LOCKING CAP, THE TIP OF THE SCREWDRIVER SHAFT FOR LOCKING CAP BROKE OFF. ONE FRAGMENT FELL INTO THE PATIENT, BUT COULD BE RECOVERED AFTER SOME TIME. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 294611 | SCRDRIVERSHAFT F/LOCKCAP F/URS | HXX | SYNTHES GMBH | 3436684 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 53 YR |