FDA Adverse Event Malfunction Summary report: N

840 VENTILATOR

MDR report key: 3193245 · Received June 25, 2013

Report

Report Number
8020893-2013-01385
Event Type
Malfunction
Date Received
June 25, 2013
Date of Event
May 10, 2013
Report Date
May 7, 2013
Manufacturer
COVIDIEN, FORMERLY NELLCOR
Product Code
CBK
PMA / PMN Number
K970460
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4).

Description of Event or Problem · 1

THE SERVICE REPORT SHOWS THE CUSTOMER REPORTED THAT THE 840 VENTILATOR STOPPED CYCLING. THERE WAS NO PT INVOLVEMENT. THE COVIDIEN CUSTOMER SUPPORT ENGINEER (CSE) VERIFIED THE MALFUNCTION. THE CSE INSPECTED THE DEVICE AND REPLACED THE O2 PSOL VALVE, EXHALATION FLOW SENSOR AND INSPIRATORY FLOW SENSOR. THE UNIT PASSED EXTENDED SELF-TESTING.