FDA Adverse Event
Malfunction
Summary report: N
840 VENTILATOR
MDR report key: 3193245
·
Received June 25, 2013
Report
- Report Number
- 8020893-2013-01385
- Event Type
- Malfunction
- Date Received
- June 25, 2013
- Date of Event
- May 10, 2013
- Report Date
- May 7, 2013
- Manufacturer
- COVIDIEN, FORMERLY NELLCOR
- Product Code
- CBK
- PMA / PMN Number
- K970460
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
(B)(4).
Description of Event or Problem · 1
THE SERVICE REPORT SHOWS THE CUSTOMER REPORTED THAT THE 840 VENTILATOR STOPPED CYCLING. THERE WAS NO PT INVOLVEMENT. THE COVIDIEN CUSTOMER SUPPORT ENGINEER (CSE) VERIFIED THE MALFUNCTION. THE CSE INSPECTED THE DEVICE AND REPLACED THE O2 PSOL VALVE, EXHALATION FLOW SENSOR AND INSPIRATORY FLOW SENSOR. THE UNIT PASSED EXTENDED SELF-TESTING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 288940 | 840 VENTILATOR | VENTILATOR | CBK | COVIDIEN, FORMERLY NELLCOR | 840 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |