HUT EXT DR FINAL ASSY-STANDARD
MDR report key: 3193193
·
Received June 25, 2013
Report
- Report Number
- 1518293-2013-00145
- Event Type
- Malfunction
- Date Received
- June 25, 2013
- Date of Event
- June 17, 2013
- Report Date
- June 25, 2013
- Manufacturer
- LIEBEL-FLARSHEIM CO.
- Product Code
- IXR
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KS, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
FIELD SERVICE ENGINEER (FSE) INVESTIGATED COMPLAINT AND FOUND THE TABLE X-RAY MONITOR DID HAVE A VIDEO DISPLAY, BUT THE SIGNAL FROM THE INFIRMED COMPUTER TO THE MONITOR HAD FAILED. FSE TROUBLESHOT THIS TO A BAD CONNECTION, REPAIRED THE CONNECTION AND THE ISSUE WAS RESOLVED. FSE COMPLETED SYSTEM CHECKOUT PER SERVICE CHECKLIST QSSRWI4.1 AND RETURNED THE UNIT TO CUSTOMER FOR FULL SERVICE.
Description of Event or Problem · 1
(B)(4): CUSTOMER REPORTS VIA PHONE THAT DURING A RETROGRADE CYSTOGRAM WITH STENT PLACEMENT, ON A FEMALE (AGE UNK), THE SYSTEM FLUORO MONITOR FAILED. PHYSICIAN USED THE CONTROL MONITOR FOR VIEWING OF FLUORO IMAGES AND PROCEDURE WAS COMPLETED WITHOUT INCIDENT. NO REPORTED INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 288647 | HUT EXT DR FINAL ASSY-STANDARD | IXR | LIEBEL-FLARSHEIM CO. | HUT EXT DR | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |