FDA Adverse Event Malfunction Summary report: N

HUT EXT DR FINAL ASSY-STANDARD

MDR report key: 3193193 · Received June 25, 2013

Report

Report Number
1518293-2013-00145
Event Type
Malfunction
Date Received
June 25, 2013
Date of Event
June 17, 2013
Report Date
June 25, 2013
Manufacturer
LIEBEL-FLARSHEIM CO.
Product Code
IXR
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KS, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

FIELD SERVICE ENGINEER (FSE) INVESTIGATED COMPLAINT AND FOUND THE TABLE X-RAY MONITOR DID HAVE A VIDEO DISPLAY, BUT THE SIGNAL FROM THE INFIRMED COMPUTER TO THE MONITOR HAD FAILED. FSE TROUBLESHOT THIS TO A BAD CONNECTION, REPAIRED THE CONNECTION AND THE ISSUE WAS RESOLVED. FSE COMPLETED SYSTEM CHECKOUT PER SERVICE CHECKLIST QSSRWI4.1 AND RETURNED THE UNIT TO CUSTOMER FOR FULL SERVICE.

Description of Event or Problem · 1

(B)(4): CUSTOMER REPORTS VIA PHONE THAT DURING A RETROGRADE CYSTOGRAM WITH STENT PLACEMENT, ON A FEMALE (AGE UNK), THE SYSTEM FLUORO MONITOR FAILED. PHYSICIAN USED THE CONTROL MONITOR FOR VIEWING OF FLUORO IMAGES AND PROCEDURE WAS COMPLETED WITHOUT INCIDENT. NO REPORTED INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
288647 HUT EXT DR FINAL ASSY-STANDARD IXR LIEBEL-FLARSHEIM CO. HUT EXT DR NA

Patients

Seq Age Sex Outcome Treatment
1 UNK