STELLARIS 25GA COMB VITRECTOMY PK W WF
Report
- Report Number
- 1920664-2013-00146
- Event Type
- Malfunction
- Date Received
- June 25, 2013
- Date of Event
- May 28, 2013
- Report Date
- May 28, 2013
- Manufacturer
- BAUSCH & LOMB, INC.
- Product Code
- HQC
- PMA / PMN Number
- K101325
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- VA, US
- Reporter Occupation
- PHYSICIAN
Narratives
EVAL COMPLETED. ONE 25GA CUTTER WAS RETURNED IN A PLASTIC ZIP LOCK BAG. THE ORIGINAL PACKAGING WAS NOT RETURNED. THE PART NUMBER AND LOT NUMBER CANNOT BE VERIFIED. THE TIP PROTECTOR WAS IN PLACE. THE NEEDLE WAS SLIGHTLY BENT APPROX 5 DEGREES OR LESS. THE PORT WINDOW WAS IN THE OPENED POSITION. THERE WAS DEBRIS VISIBLE INSIDE THE PORT AND AROUND THE PORT WINDOW ON THE OUTER NEEDLE SHAFT. THERE WAS DRIED SOLUTION IN THE TUBING. THE EXTENSION HANDLE WAS AROUND THE TUBING BUT NO ATTACHED TO THE CUTTER BACK CAP. THE BACK CAP WAS ALIGNED CORRECTLY WITH THE VENT HOLE IN THE SIDE OF THE CUTTER BODY. A FUNCTIONAL TEST WAS PERFORMED USING A STELLARIS PC SYSTEM. THE CUTTER DID RELEASE ONE AIR BUBBLE FROM THE PORT AT THE INITIAL START UP. THE CUTTER HAD GOOD CLEAN CUTS AT VARIOUS CUT RATES. THE CUTTER HAD GOOD ASPIRATION AND FUNCTIONED AS REQUIRED. THE STERILIZATION AND LOT HISTORY RECORDS HAVE BEEN REVIEWED AND FOUND TO BE ACCEPTABLE.
THE USER FACILITY REPORTED VITRECTOMY CUTTER WAS NOT PERFORMING UPON ACTIVATION. THERE WERE BUBBLES COMING FROM THE TOP OF THE CUTTER AND IT WAS MAKING A NOISE. THE CUTTER SUBSEQUENTLY BEGAN TO CUT PROPERLY AND THE PROCEDURE WAS COMPLETELY WITH NO FURTHER INCIDENTS. THERE WAS PT CONTACT BUT NO INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 288178 | STELLARIS 25GA COMB VITRECTOMY PK W WF | HQC | BAUSCH & LOMB, INC. | BL5325W | V0289 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |