ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2013-09123
- Event Type
- Malfunction
- Date Received
- June 27, 2013
- Report Date
- May 31, 2013
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- OTHER
Narratives
THE PUMP HAS BEEN RETURNED TO ANIMAS. EVALUATION HAS NOT YET BEEN COMPLETED. WHEN EVALUATION IS COMPLETE A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSION CAN BE MADE AT THIS TIME.
FOLLOW-UP #1 DATE OF SUBMISSION 08/21/2012, DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2012 WITH THE FOLLOWING FINDINGS: THE KEYPAD WAS FOUND TO BE INTACT. THE KEYPAD BUTTONS WERE FOUND TO BE RESPONSIVE. THE KEYPAD COVER WAS REMOVED AND CONTAMINATION WAS FOUND UNDER THE KEY CONTACTS.
ON (B)(6) 2013, THE REPORTER CONTACTED ANIMAS, ALLEGING A BUTTON/KEYPAD (TACTILE CHANGES/UNRESPONSIVE) ISSUE. THE REPORTER STATED THAT THE UP ARROW, DOWN ARROW, AND OK KEYPAD BUTTONS WERE INTERMITTENTLY UNRESPONSIVE AND REQUIRED MULTIPLE, HARD PRESSES TO ELICIT A RESPONSE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 294693 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 19 YR |