FDA Adverse Event
Malfunction
Summary report: N
CORE IMPACTION DRILL
MDR report key: 3192945
·
Received June 27, 2013
Report
- Report Number
- 0001811755-2013-01489
- Event Type
- Malfunction
- Date Received
- June 27, 2013
- Date of Event
- June 10, 2013
- Report Date
- June 10, 2013
- Manufacturer
- STRYKER INSTRUMENTS-KALAMAZOO
- Product Code
- DZJ
- PMA / PMN Number
- K040369
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
UPON DISASSEMBLY FOR VISUAL INSPECTION CORROSION WAS FOUND ON THE NOSE CONE. THE ROTOR IS STUCK IN THE MOTOR, THE SPINDLE HOUSING IS WORN AND THE DRIVESHAFT BEARINGS ARE GRITTY.
Description of Event or Problem · 1
WHILE TESTING THE CORE IMPACTION DRILL AT THE MANUFACTURER IT WAS REPORTED THAT THE DEVICE HEATED UP. THERE WAS NO PATIENT INVOLVEMENT OR ADVERSE CONSEQUENCES REPORTED WITH THIS EVENT.