ANIMAS INSULIN INFUSION PUMP
Report
- Report Number
- 2531779-2013-09104
- Event Type
- Malfunction
- Date Received
- June 27, 2013
- Report Date
- May 30, 2013
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K042873
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- PATIENT
Narratives
THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2013 WITH THE FOLLOWING FINDINGS: THE PUMP'S VIBRATORY FEATURE WAS FOUND NOT TO BE FUNCTIONING. THE AUDIBLE FEATURE WAS FOUND TO BE WORKING APPROPRIATELY. THE PUMP WAS OPENED AND THE VIBRATION MOTOR CONNECTOR PINS WERE FOUND TO BE MISALIGNED. THE VIBRATION MOTOR WAS CONNECTED TO AN EXTERNAL POWER SOURCE AND THE VIBRATORY FEATURE WAS FOUND TO BE WORKING PROPERLY.
ON (B)(6) 2013, THE REPORTER CONTACTED ANIMAS, ALLEGING A AUDIO TONE/VIBRATION (NO VIBRATION) ISSUE. THE PUMP WAS RETURNED AND INVESTIGATED ON (B)(4) 2013. THE PUMP'S VIBRATORY FEATURE WAS FOUND NOT TO BE FUNCTIONING. THE AUDIBLE FEATURE WAS FOUND TO BE WORKING APPROPRIATELY. THE PUMP WAS OPENED AND THE VIBRATION MOTOR CONNECTOR PINS WERE FOUND TO BE MISALIGNED. THIS COMPLAINT IS BEING REPORTED DUE TO THE VIBRATORY FEATURE NOT WORKING ON THE PUMP. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 294777 | ANIMAS INSULIN INFUSION PUMP | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 15 YR |