FDA Adverse Event Malfunction Summary report: N

ANIMAS INSULIN INFUSION PUMP

MDR report key: 3192876 · Received June 27, 2013

Report

Report Number
2531779-2013-09104
Event Type
Malfunction
Date Received
June 27, 2013
Report Date
May 30, 2013
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K042873
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2013 WITH THE FOLLOWING FINDINGS: THE PUMP'S VIBRATORY FEATURE WAS FOUND NOT TO BE FUNCTIONING. THE AUDIBLE FEATURE WAS FOUND TO BE WORKING APPROPRIATELY. THE PUMP WAS OPENED AND THE VIBRATION MOTOR CONNECTOR PINS WERE FOUND TO BE MISALIGNED. THE VIBRATION MOTOR WAS CONNECTED TO AN EXTERNAL POWER SOURCE AND THE VIBRATORY FEATURE WAS FOUND TO BE WORKING PROPERLY.

Description of Event or Problem · 1

ON (B)(6) 2013, THE REPORTER CONTACTED ANIMAS, ALLEGING A AUDIO TONE/VIBRATION (NO VIBRATION) ISSUE. THE PUMP WAS RETURNED AND INVESTIGATED ON (B)(4) 2013. THE PUMP'S VIBRATORY FEATURE WAS FOUND NOT TO BE FUNCTIONING. THE AUDIBLE FEATURE WAS FOUND TO BE WORKING APPROPRIATELY. THE PUMP WAS OPENED AND THE VIBRATION MOTOR CONNECTOR PINS WERE FOUND TO BE MISALIGNED. THIS COMPLAINT IS BEING REPORTED DUE TO THE VIBRATORY FEATURE NOT WORKING ON THE PUMP. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
294777 ANIMAS INSULIN INFUSION PUMP INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 15 YR