FDA Adverse Event
Malfunction
Summary report: N
PARADIGM REAL-TIME INSULIN INFUSION PUMP
MDR report key: 3192818
·
Received June 27, 2013
Report
- Report Number
- 3004209178-2013-95120
- Event Type
- Malfunction
- Date Received
- June 27, 2013
- Date of Event
- June 9, 2013
- Report Date
- June 9, 2013
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MO
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
MOTOR ERROR ALARM DURING THE BASIC OCCLUSION TEST DUE TO PROTRUDED/LOOSE DRIVE SUPPORT DISK. MOTOR PASSED MOTOR TEST. NO PRIME ALARM.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE INSULIN PUMP ALARMED DURING PRIME. THE DRIVE SUPPORT CAP IS PROTRUDED. ADVISED CUSTOMER TO DISCONTINUE USE OF THE INSULIN PUMP. ADVISED CUSTOMER THAT THE INSULIN PUMP WILL BE REPLACED. NOTHING FURTHER REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 293981 | PARADIGM REAL-TIME INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-722NAS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 39 YR |