FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 3192778 · Received June 27, 2013

Report

Report Number
3004209178-2013-95106
Event Type
Malfunction
Date Received
June 27, 2013
Date of Event
June 8, 2013
Report Date
June 8, 2013
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP ALARMED PRIME ALARM DURING THE PRIME TEST DUE TO PROTRUDED/LOOSE DRIVE SUPPORT DISK. THE INSULIN PUMP HAD A MISSING END CAP STICKER.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULIN PUMP HAD TROUBLE PRIMING. CUSTOMER RECEIVED DURING PRIME. CUSTOMER'S BLOOD GLUCOSE READING IS 109 MG/DL. THE DRIVE SUPPORT CAP IS LOOSE. NOTHING FURTHER REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
291918 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-522NAS

Patients

Seq Age Sex Outcome Treatment
1 69 YR Life Threatening