FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 3192773 · Received June 27, 2013

Report

Report Number
3004209178-2013-95101
Event Type
Malfunction
Date Received
June 27, 2013
Date of Event
June 8, 2013
Report Date
June 8, 2013
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PATIENT FAMILY MEMBER OR FRIEND

Narratives

Additional Manufacturer Narrative · 1

THE INFUSION PUMP ALARM DURING THE PRIME TEST DUE TO PROTRUDED/LOOSE DRIVE SUPPORT DISK.

Description of Event or Problem · 1

IT WAS REPORTED BY CUSTOMER'S SPOUSE, THAT THE INSULIN PUMP ALARMED, AND INSULIN SQUIRTED OUT DURING THE MANUAL PRIME PROCESS. CUSTOMER'S BLOOD GLUCOSE READING IS 90 MG/DL. THE DRIVE SUPPORT CAP IS PROTRUDED. NOTHING FURTHER REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
291806 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-722NAS

Patients

Seq Age Sex Outcome Treatment
1 58 YR