COULTER® ACT DIFF 2? ANALYZER
Report
- Report Number
- 1061932-2013-01291
- Event Type
- Malfunction
- Date Received
- June 27, 2013
- Date of Event
- May 27, 2013
- Report Date
- May 27, 2013
- Manufacturer
- BECKMAN COULTER
- Product Code
- GKZ
- PMA / PMN Number
- K990352
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
SERVICE WAS NOT DISPATCHED AS THE CUSTOMER RESOLVED THE ISSUE THROUGH SYSTEM TROUBLESHOOTING. ALL RELATED MEDWATCH REPORTS ASSOCIATED WITH THIS INCIDENT: 1061932-2013-01260, 1061932-2013-01290, 1061932-2013-01291.
THE AFFILIATE REPORTED THE CUSTOMER ALLEGED WHITE BLOOD CELL (WBC) FLAG AND VOTE-OUTS FOR TWO PATIENTS, AND ELEVATED INTERMITTENT PLATELET BACKGROUND FOR ONE PATIENT, INVOLVING THE COULTER ACT DIFF 2 ANALYZER. RESULTS FOR THREE PATIENTS WERE PROVIDED SHOWING ERRONEOUS WBC AND DIFFERENTIAL RESULT WITH INSTRUMENT GENERATED MESSAGES. ALL OTHER PARAMETERS WERE COMPARABLE. MAINTENANCE LOG INDICATED SYSTEM STARTUP WAS REPEATED MULTIPLE TIMES ON VARIOUS DAYS DUE TO ELEVATED PLATELET BACKGROUND. NO ERRONEOUS RESULTS WERE RELEASED FROM THE LABORATORY. THERE WAS NO PATIENT CONSEQUENCE OR CHANGE IN PATIENT TREATMENT ASSOCIATED WITH THIS EVENT. THE CUSTOMER REPLACED THE CHECK VALVES AT THE WBC BATH AND THE RBC (RED BLOOD BLOOD) PORT AS INSTRUCTED BY THE FIELD SERVICE ENGINEER (FSE) AND RESOLVED THE ISSUE. THE INSTRUMENT WAS RETURNED TO NORMAL OPERATION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 292699 | COULTER® ACT DIFF 2? ANALYZER | COUNTER, DIFFERENTIAL CELL | GKZ | BECKMAN COULTER | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 92 YR |