PARADIGM REAL-TIME INSULIN INFUSION PUMP
MDR report key: 3192661
·
Received June 27, 2013
Report
- Report Number
- 2032227-2013-02596
- Event Type
- Malfunction
- Date Received
- June 27, 2013
- Date of Event
- June 17, 2013
- Report Date
- June 17, 2013
- Manufacturer
- MEDTRONIC MINIMED
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
THE INSULIN PUMP ALARMED DURING THE BASIC OCCLUSION TEST, AND WAS UNABLE TO PRIME DURING THE PRIME TEST DUE TO A LOOSE DRIVE SUPPORT DISK. THE INSULIN PUMP HAD A MISSING END CAP STICKER.
Description of Event or Problem · 1
THE CUSTOMER STATED THAT THE INSULIN PUMP GAVE AN ALARM DURING THE MANUAL PRIME. THE CUSTOMER STATED THAT SHE GOT THE ALARM MULTIPLE TIMES. ADVISED THE CUSTOMER THAT THE INSULIN PUMP WOULD BE REPLACED. NOTHING FURTHER WAS REPORTED.