FDA Adverse Event Malfunction Summary report: N

PARADIGM INSULIN INFUSION PUMP

MDR report key: 3192658 · Received June 27, 2013

Report

Report Number
2032227-2013-02591
Event Type
Malfunction
Date Received
June 27, 2013
Date of Event
June 14, 2013
Report Date
June 14, 2013
Manufacturer
MEDTRONIC MINIMED
Product Code
LZG
PMA / PMN Number
K031390
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP ALARMED MOTOR ERROR DURING THE BASIC OCCLUSION TEST DUE TO A PROTRUDED/LOOSE DRIVE SUPPORT DISK. THE INSULIN PUMP HAD A MISSING END CAP STICKER, AND WAS RECEIVED WITH CRACKED WINDOW CORNERS. THE MOTOR PASSED THE MOTOR TEST.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULIN PUMP HAD A CRACK NEAR THE DISPLAY. IT WAS NOT KNOWN HOW THE CRACK OCCURRED. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
292620 PARADIGM INSULIN INFUSION PUMP INSULIN INFUSION PUMP LZG MEDTRONIC MINIMED MMT-712LWWL

Patients

Seq Age Sex Outcome Treatment
1