FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 3192635 · Received June 27, 2013

Report

Report Number
2032227-2013-02585
Event Type
Malfunction
Date Received
June 27, 2013
Date of Event
June 2, 2013
Report Date
June 2, 2013
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP ALARMED MOTOR ERROR DURING THE BASIC OCCLUSION TEST DUE TO A PROTRUDED/LOOSE DRIVE SUPPORT DISK. THE MOTOR PASSED THE MOTOR TEST. THE INSULIN PUMP PASSED THE DISPLACEMENT TEST.

Description of Event or Problem · 1

THE CUSTOMER STATED THAT THE INSULIN PUMP GAVE A MOTOR ERROR ALARM WHEN HE ATTEMPTED TO TREAT A HIGH BLOOD GLUCOSE OF 368 MG/DL. TROUBLESHOOTING WAS PERFORMED, AND THE INSULIN PUMP ALARMED MOTOR ERROR DURING THE DISPLACEMENT TEST. THE CUSTOMER INSPECTED THE DRIVE SUPPORT CAP, AND IT DID NOT APPEAR DAMAGED. ADVISED THE CUSTOMER THAT THE INSULIN PUMP WOULD BE REPLACED. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
292546 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-522LNAB

Patients

Seq Age Sex Outcome Treatment
1 45 YR