FDA Adverse Event
Malfunction
Summary report: N
PARADIGM REAL-TIME INSULIN INFUSION PUMP
MDR report key: 3192635
·
Received June 27, 2013
Report
- Report Number
- 2032227-2013-02585
- Event Type
- Malfunction
- Date Received
- June 27, 2013
- Date of Event
- June 2, 2013
- Report Date
- June 2, 2013
- Manufacturer
- MEDTRONIC MINIMED
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
THE INSULIN PUMP ALARMED MOTOR ERROR DURING THE BASIC OCCLUSION TEST DUE TO A PROTRUDED/LOOSE DRIVE SUPPORT DISK. THE MOTOR PASSED THE MOTOR TEST. THE INSULIN PUMP PASSED THE DISPLACEMENT TEST.
Description of Event or Problem · 1
THE CUSTOMER STATED THAT THE INSULIN PUMP GAVE A MOTOR ERROR ALARM WHEN HE ATTEMPTED TO TREAT A HIGH BLOOD GLUCOSE OF 368 MG/DL. TROUBLESHOOTING WAS PERFORMED, AND THE INSULIN PUMP ALARMED MOTOR ERROR DURING THE DISPLACEMENT TEST. THE CUSTOMER INSPECTED THE DRIVE SUPPORT CAP, AND IT DID NOT APPEAR DAMAGED. ADVISED THE CUSTOMER THAT THE INSULIN PUMP WOULD BE REPLACED. NOTHING FURTHER WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 292546 | PARADIGM REAL-TIME INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC MINIMED | MMT-522LNAB |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 45 YR |