FDA Adverse Event
Malfunction
Summary report: N
NEPTUNE ROVER WITH SMOKE EVAC AND POWER POLE
MDR report key: 3192478
·
Received June 27, 2013
Report
- Report Number
- 0001811755-2013-01480
- Event Type
- Malfunction
- Date Received
- June 27, 2013
- Date of Event
- June 3, 2013
- Report Date
- June 3, 2013
- Manufacturer
- STRYKER INSTRUMENTS-KALAMAZOO
- Product Code
- FYD
- PMA / PMN Number
- K012991
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE MANUFACTURER SERVICE TECHNICIAN REPAIRED THE POWER CORD AND RETURNED THE UNIT TO SERVICE.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE ROVER HAD A FRAYED POWER CORD. WHEN THE FIELD SERVICE TECHNICIAN EVALUATED THE UNIT HE REPORTED THAT THE PLUG WAS LOOSE AND HAD EXPOSED INNER WIRING. THERE WAS NO PATIENT INVOLVEMENT AND NO ADVERSE CONSEQUENCES ASSOCIATED WITH THE DEVICE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 294841 | NEPTUNE ROVER WITH SMOKE EVAC AND POWER POLE | APPARATUS, EXHAUST, SURGICAL | FYD | STRYKER INSTRUMENTS-KALAMAZOO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |