FDA Adverse Event Injury Summary report: N

PROMUS ELEMENT? PLUS

MDR report key: 3192475 · Received June 27, 2013

Report

Report Number
2134265-2013-04219
Event Type
Injury
Date Received
June 27, 2013
Date of Event
May 29, 2013
Report Date
May 30, 2013
Manufacturer
BOSTON SCIENTIFIC - GALWAY
Product Code
NIQ
PMA / PMN Number
P110010
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WI, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

DEVICE IS A COMBINATION PRODUCT. (B)(4). DEVICE EVALUATED BY MANUFACTURER: IT IS INDICATED THAT THE DEVICE WILL NOT BE RETURNED FOR EVALUATION; THEREFORE A FAILURE ANALYSIS OF THE COMPLAINT DEVICE COULD NOT BE COMPLETED. A REVIEW OF THE BATCH HISTORY, HISTORICAL TRENDING, AND SIMILAR COMPLAINT TRENDING REVIEW FOR THE PRODUCT FAMILY WILL BE CONDUCTED. IF THERE IS ANY FURTHER RELEVANT INFORMATION FROM THAT REVIEW, A SUPPLEMENTAL MEDWATCH WILL BE FILED. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A PERCUTANEOUS CORONARY INTERVENTION, STENT DAMAGE OCCURRED. THE TARGET LESION WAS LOCATED IN THE MILDLY TORTUOUS AND MILDLY CALCIFIED PROXIMAL LEFT ANTERIOR DESCENDING ARTERY (LAD). A 3.50MM X 16MM PROMUS ELEMENT PLUS STENT WAS IMPLANTED AT 11 ATMS IN THE TARGET LESION. UPON POST-DILATION USING A 4.00MM X 12MM NON-BSC BALLOON CATHETER, THE PROXIMAL PORTION OF THE IMPLANTED STENT WAS COMPRESSED LONGITUDINALLY AT "APPROXIMATELY 4 TO 6MMS." A 4.00MM X 12MM PROMUS ELEMENT PLUS STENT WAS THEN PLACED OVER THE LONGITUDINALLY COMPRESSED AREA OF THE FIRST IMPLANTED STENT AND EXTENDED PROXIMALLY IN THE PROXIMAL LAD. THE PROCEDURE WAS COMPLETED AND NO PATIENT COMPLICATIONS WERE REPORTED. THE PATIENT STATUS WAS FINE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
294840 PROMUS ELEMENT? PLUS STENT, CORONARY, DRUG-ELUTING NIQ BOSTON SCIENTIFIC - GALWAY H7493911416350 16028898

Patients

Seq Age Sex Outcome Treatment
1 56 YR Required Intervention BALLOON CATHETER: .00 X 12MM MEDTRONIC NC SPRINTER