FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 3192448 · Received June 27, 2013

Report

Report Number
2531779-2013-09076
Event Type
Malfunction
Date Received
June 27, 2013
Report Date
June 24, 2013
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Additional Manufacturer Narrative · 1

FOLLOW-UP #1: DATE OF SUBMISSION 10/29/2013. DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 10/18/2013 WITH THE FOLLOWING FINDINGS: VISUAL INSPECTION SHOWS EVIDENCE OF MOISTURE CORROSION ON THE DISPLAY SCREEN INSIDE THE PUMP. THE PUMP IS UNABLE TO POWER ¿UP¿ AND TO DISPLAY ¿VERIFY¿ SCREEN. UNABLE TO DOWNLOAD THE BLACK BOX AND THE HISTORY DATA. INVESTIGATORS WERE UNABLE TO ADEQUATELY INVESTIGATE REPORTED ¿TEMPERATURE ISSUE¿.

Description of Event or Problem · 1

THE REPORTER CONTACTED ANIMAS ON (B)(6) 2013 REPORTING THAT THE PUMP WAS IN THE WATER TODAY AND IS NOW HOT. THE PATIENT STATED THAT THE BATTERY AND PUMP ARE HOT. THERE IS NO REPORTED ADVERSE EVENT WITH THIS COMPLAINT. THIS REPORT IS MADE BASED ON THE ALLEGATION OF THE TEMPERATURE ISSUE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
294836 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 18 YR