FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 3192443 · Received June 27, 2013

Report

Report Number
2531779-2013-09077
Event Type
Malfunction
Date Received
June 27, 2013
Report Date
May 31, 2013
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS BEEN RETURNED TO ANIMAS. EVALUATION HAS NOT YET BEEN COMPLETED. WHEN EVALUATION IS COMPLETE A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSION CAN BE MADE AT THIS TIME.

Additional Manufacturer Narrative · 1

FOLLOW-UP # 1 (B)(4) DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2013 WITH THE FOLLOWING FINDINGS: DURING INVESTIGATION, THE COMPLAINT OF NO SOUND FROM THE ALARM WAS NOT DUPLICATED; TESTING REVEALED THAT THE AUDIBLE ALARM WAS WITHIN SPECIFICATIONS AND WAS FUNCTIONAL. UNRELATED TO THE ALARM ISSUE, EVALUATION REVEALED A FADED, DISCOLORED DISPLAY THAT WAS DIFFICULT TO READ. A TEST SCREEN WAS INSERTED AND WAS FOUND TO FUNCTION PROPERLY WITH NO VISIBLE SIGNS OF FADING OR DISCOLORATION. THE PUMP CASE WAS REMOVED, AND NO EVIDENCE OF MOISTURE CONTAMINATION OR LOOSE ELECTRICAL COMPONENTS WAS IDENTIFIED INSIDE PUMP.

Description of Event or Problem · 1

ON (B)(6) 2013, THE REPORTER CONTACTED ANIMAS, ALLEGING AN (AUDIO TONE/VIBRATION (NO SOUNDS)) ISSUE. IT WAS REPORTED THAT THE PUMP HAD NO AUDIBLE TONES. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
293368 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 47 YR