FDA Adverse Event Injury Summary report: N

NEUROFORM 3 EZ- 4.5 X 30MM

MDR report key: 3192429 · Received June 27, 2013

Report

Report Number
3008853977-2013-00202
Event Type
Injury
Date Received
June 27, 2013
Date of Event
May 18, 2013
Report Date
May 31, 2013
Manufacturer
STRYKER NEUROVASCULAR CORK
Product Code
NJE
PMA / PMN Number
H020002
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

SUBJECT DEVICE REMAINS IMPLANTED.

Additional Manufacturer Narrative · 1

THE SUBJECT DEVICE IS NOT AVAILABLE; THEREFORE ANALYSIS CANNOT BE PERFORMED. FROM THE INFORMATION AVAILABLE, THERE IS NO INDICATION THAT THE REPORTED EVENT IS RELATED TO PRODUCT SPECIFICATIONS NONCONFORMANCE OR MISUSE. THERE IS ALSO NO INDICATION THAT THE SUBJECT DEVICE MALFUNCTIONED. HOWEVER, PATIENT COMPLICATIONS IS A KNOWN RISK ASSOCIATED WITH SUCH PROCEDURES AND NOTED AS SUCH IN THE DIRECTIONS FOR USE (DFU). THEREFORE, A ROOT CAUSE OF ANTICIPATED PROCEDURAL COMPLICATION HAS BEEN ASSIGNED TO THE EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT UNDERWENT A PROCEDURE TO TREAT AN ANEURYSM LOCATED IN THE LEFT INTERNAL CAROTID ARTERY. THREE STENTS WERE USED TO COVER THE NECK OF THE ANEURYSM. WHEN THE PHYSICIAN APPROACHED THE ANEURYSM WITH THE MICROCATHETER TO DEPLOY A FOURTH STENT, THE CATHETER PUSHED THE PROXIMAL END OF THE THIRD STENT. THE PROXIMAL END OF THE THIRD STENT MIGRATED INTO THE ANEURYSM. AS A RESULT, THE PHYSICIAN CHANGED THE TREATMENT STRATEGY FROM STENT ASSISTED TECHNIQUE TO USING COILS TO OCCLUDE THE INTERNAL CAROTID ARTERY. AFTER THE PROCEDURE, THE PATIENT EXPERIENCED RIGHT SIDE PARALYSIS AND APHASIA. NO FURTHER INFORMATION WAS PROVIDED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT UNDERWENT A PROCEDURE TO TREAT AN ANEURYSM LOCATED IN THE LEFT INTERNAL CAROTID ARTERY. THREE STENTS WERE USED TO COVER THE NECK OF THE ANEURYSM. WHEN THE PHYSICIAN APPROACHED THE ANEURYSM WITH THE MICROCATHETER TO DEPLOY A FOURTH STENT, THE CATHETER PUSHED THE PROXIMAL END OF THE THIRD STENT. THE PROXIMAL END OF THE THIRD STENT MIGRATED INTO THE ANEURYSM. AS A RESULT, THE PHYSICIAN CHANGED THE TREATMENT STRATEGY FROM STENT ASSISTED TECHNIQUE TO USING COILS TO OCCLUDE THE INTERNAL CAROTID ARTERY. AFTER THE PROCEDURE, THE PATIENT EXPERIENCED RIGHT SIDE PARALYSIS AND APHASIA. NO FURTHER INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
294725 NEUROFORM 3 EZ- 4.5 X 30MM STENT, INTRACRANIAL, NEUROVASCULAR NJE STRYKER NEUROVASCULAR CORK 15877245

Patients

Seq Age Sex Outcome Treatment
1 Other 29 COILS (MANUFACTURER UNKNOWN)| 2 NEUROFORM EZ STENTS (STRYKER)| 4 GDC COILS (STRYKER)