FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 3192385 · Received June 27, 2013

Report

Report Number
2531779-2013-09073
Event Type
Malfunction
Date Received
June 27, 2013
Report Date
June 1, 2013
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

FOLLOW-UP #1 (B)(4) DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2013 WITH THE FOLLOWING FINDINGS: A REVIEW OF THE BLACK BOX SHOWED NO CARTRIDGE DETECTED AND MULTIPLE LOSS OF PRIMES DUE TO ZERO FORCE WARNINGS. THE PUMP POWERS ON AND DISPLAYS VERIFY SCREEN. THE DISPLAY SCREEN IS DIM AND REDDISH. A TEST DISPLAY WAS MOUNTED DURING INVESTIGATION AND SCREEN FULLY ILLUMINATED. THE PUMP PASSES EZ PRIME STEPS AND PRIMES CORRECTLY. INVESTIGATORS WERE UNABLE TO DUPLICATE REPORTED LOAD STEP MALFUNCTION. FORCE SENSOR CALIBRATION READING IS AT THE HIGHEST COUNT. PUMP¿S COVER WAS REMOVED, MOISTURE CORROSION WAS OBSERVED INSIDE THE UNIT FOUR-THE FORCE SENSOR CIRCUIT WAS DISASSEMBLED, MOISTURE CORROSION WAS FOUND TO THE FORCE SENSOR PLATE. FORCE SENSOR RESISTANCE READING IS WITHIN SPECIFICATIONS.

Additional Manufacturer Narrative · 1

THE PUMP HAS BEEN RETURNED TO ANIMAS. EVALUATION HAS NOT YET BEEN COMPLETED. WHEN EVALUATION IS COMPLETE A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSION CAN BE MADE AT THIS TIME.

Description of Event or Problem · 1

ON (B)(6) 2013, THE REPORTER CONTACTED ANIMAS, ALLEGING A PRIME (LOAD STEP MALFUNCTION) ISSUE. THE REPORTER STATED THAT THE PUMP EMPTIED THE CARTRIDGE OF INSULIN DURING THE LOAD STEP. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
292633 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 19 YR