FDA Adverse Event Death Summary report: N

CONSULTA CRT-D

MDR report key: 3192375 · Received June 27, 2013

Report

Report Number
3004209178-2013-10954
Event Type
Death
Date Received
June 27, 2013
Date of Event
May 30, 2013
Report Date
May 30, 2013
Manufacturer
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
Product Code
NIK
PMA / PMN Number
P010031
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. ATTEMPTS TO OBTAIN ADDITIONAL INFORMATION WERE UNSUCCESSFUL. CONCOMITANT PRODUCTS: 419688 IMPLANTABLE PACING LEAD IMPLANTED: 2010 (B)(6); 693565 IMPLANTABLE TACHY LEAD IMPLANTED: 2010 (B)(6); 507 6-45 IMPLANTABLE PACING LEAD IMPLANTED: 2007 (B)(6). (B)(4).

Description of Event or Problem · 1

A CALL WAS RECEIVED FROM THE CLINICIAN REQUESTING REVIEW OF A REMOTE MONITOR TRANSMISSION. THE TRANSMISSION OCCURRED DUE TO TRIGGERING OF A LEAD INTEGRITY ALERT (LIA). DURING THE CONVERSATION IT WAS REPORTED THAT THE PATIENT IS DECEASED. TECHNICAL SERVICE REVIEW OF THE TRANSMISSION WITH THE CLINICIAN NOTED THAT THE PATIENT HAD VENTRICULAR FIBRILLATION (VF) AND RECEIVED SIX SHOCKS WITHOUT SUCCESSFULLY TERMINATING THE RHYTHM. THE LIA IS NOTED TO BE DUE TO NONPHYSIOLOGIC SENSING AND ELEVATED SHORT INTERVAL COUNTS (SIC). ADDITIONAL INFORMATION RELATED TO THE CAUSE OF DEATH AND CIRCUMSTANCES OF DEATH HAVE BEEN REQUESTED AND NOT RECEIVED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
294620 CONSULTA CRT-D DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO NIK MEDTRONIC MED REL MEDTRONIC PUERTO RICO D224TRK

Patients

Seq Age Sex Outcome Treatment
1 00061 YR Death