FDA Adverse Event Malfunction Summary report: N

LIFEVEST WCD 4000 SYSTEM

MDR report key: 3192355 · Received June 18, 2013

Report

Report Number
3008642652-2013-01601
Event Type
Malfunction
Date Received
June 18, 2013
Date of Event
May 21, 2013
Report Date
June 12, 2013
Manufacturer
ZOLL LIFECOR
Product Code
MVK
PMA / PMN Number
P010030
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF ELECTRODE BELT SN (B)(4) HAS BEEN COMPLETED. UPON EVALUATION, THE CABLE CONNECTING THE REAR THERAPY ELECTRODES (TE) AND THE DISTRIBUTION NODE(DN) WAS PULLED FORM THE STRAIN RELIEF AT THE DN. THE CAUSE OF THE PULSE LEAD HI-POT FAILURE IS THE PULLED CABLE. THE ROOT CAUSE OF THE DAMAGED CABLE AND INTERNAL WIRES CANNOT BE POSITIVELY IDENTIFIED BUT IS LIKELY EXCESSIVE FORCE PLACED ON THE CABLE. NO ADVERSE EVENT RESULTED FROM THE DAMAGED CABLE AND WIRES. THE LAST PT TO USE THIS BELT DID NOT REPORT ANY DEFICIENCIES.

Description of Event or Problem · 1

REVIEW OF SERVICE DATA DETECTED A REPORTABLE EVENT. DURING SERVICING, BELT SN (B)(4) FAILED THE PULSE LEAD HI-POT TEST. THE LAST PT TO USE THIS BELT DID NOT REPORT ANY DEFICIENCIES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
276165 LIFEVEST WCD 4000 SYSTEM WEARABLE CARDIOVERTER DEFIBRILLATOR MVK ZOLL LIFECOR WCD 4000 NA

Patients

Seq Age Sex Outcome Treatment
1 NA