FDA Adverse Event Malfunction Summary report: N

9900

MDR report key: 3192246 · Received June 20, 2013

Report

Report Number
1720753-2013-07388
Event Type
Malfunction
Date Received
June 20, 2013
Date of Event
June 6, 2013
Report Date
June 20, 2013
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SVC REP PERFORMED AN ON SITE INVESTIGATION. THE REPORTED ISSUE COULD NOT BE DUPLICATED. THE CONNECTORS WERE RESEATED DURING THE SVC CALL. THE SYS WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SVC.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYS WAS STUCK IN SUBTRACTION MODE AND WERE UNABLE TO REBOOT. THIS ISSUE MAY REFER TO A LOSS OF SYS FUNCTIONALITY. THERE IS NO REPORT OF PT INJURY OR DEATH.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
281534 9900 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9900

Patients

Seq Age Sex Outcome Treatment
1