Narratives
Additional Manufacturer Narrative · 1
DURING THE EVAL OF THE DEVICE AT THE MFR'S SERVICE CENTER, A "SERVICE REQUIRED" CODE WAS FOUND IN THE VENTILATOR'S DOWNLOADED ERROR LOG. THE DEVICE'S SYSTEM BOARD, POWER MANAGEMENT BOARD, AND INTERNAL BATTERY WERE REPLACED TO ADDRESS THE ISSUE. THIS REPORT IS BEING SUBMITTED AS PART OF A PROGRAM TO RETROSPECTIVELY REVIEW POSSIBLE DEVICE MALFUNCTIONS THAT DID NOT INVOLVE ANY INJURY, TO DETERMINE WHETHER THEY ARE REPORTABLE UNDER 21 CFR PART 803 AND THE MFR'S UPDATED REPORTING PROCEDURES. THIS PROGRAM IS BEING UNDERTAKEN TO ADDRESS FDA WARNING LETTER 12-PHI-01.
Description of Event or Problem · 1
A VENTILATOR WAS RETURNED TO THE MFR FOR SERVICE. THE DEVICE WAS NOT IN PT USE.