FDA Adverse Event
Malfunction
Summary report: N
GARBIN 1
MDR report key: 3192186
·
Received June 20, 2013
Report
- Report Number
- 2518422-2013-01188
- Event Type
- Malfunction
- Date Received
- June 20, 2013
- Date of Event
- January 13, 2011
- Report Date
- January 13, 2011
- Manufacturer
- RESPIRONICS INC
- Product Code
- CBK
- PMA / PMN Number
- K083526
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
DURING EVAL OF THE DEVICE AT THE MFR'S SERVICE CENTER, "VENTILATOR INOPERATIVE" CODES WERE FOUND IN THE VENTILATOR'S DOWNLOADED ERROR LOG. THE DEVICE'S SENSOR BOARD, AND POWER MANAGEMENT BOARD WERE REPLACED TO ADDRESS THE ISSUE. THIS REPORT IS BEING SUBMITTED AS PART OF A PROGRAM TO RETROSPECTIVELY REVIEW POSSIBLE DEVICE MALFUNCTIONS THAT DID NOT INVOLVE ANY INJURY, TO DETERMINE WHETHER THEY ARE REPORTABLE UNDER 21 CFR PART 803 AND THE MFR'S UPDATED REPORTING PROCEDURES. THIS PROGRAM IS BEING UNDERTAKEN TO ADDRESS FDA WARNING LETTER 12-PHI-01.
Description of Event or Problem · 1
A VENTILATOR WAS RETURNED TO THE MFR FOR SERVICE. THERE WAS NO HARM OR INJURY REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 281719 | GARBIN 1 | VENTILATOR, CONTINUOUS, FACILITY USE | CBK | RESPIRONICS INC | 1058180 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|