FDA Adverse Event Malfunction Summary report: N

GARBIN 1

MDR report key: 3192186 · Received June 20, 2013

Report

Report Number
2518422-2013-01188
Event Type
Malfunction
Date Received
June 20, 2013
Date of Event
January 13, 2011
Report Date
January 13, 2011
Manufacturer
RESPIRONICS INC
Product Code
CBK
PMA / PMN Number
K083526
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DURING EVAL OF THE DEVICE AT THE MFR'S SERVICE CENTER, "VENTILATOR INOPERATIVE" CODES WERE FOUND IN THE VENTILATOR'S DOWNLOADED ERROR LOG. THE DEVICE'S SENSOR BOARD, AND POWER MANAGEMENT BOARD WERE REPLACED TO ADDRESS THE ISSUE. THIS REPORT IS BEING SUBMITTED AS PART OF A PROGRAM TO RETROSPECTIVELY REVIEW POSSIBLE DEVICE MALFUNCTIONS THAT DID NOT INVOLVE ANY INJURY, TO DETERMINE WHETHER THEY ARE REPORTABLE UNDER 21 CFR PART 803 AND THE MFR'S UPDATED REPORTING PROCEDURES. THIS PROGRAM IS BEING UNDERTAKEN TO ADDRESS FDA WARNING LETTER 12-PHI-01.

Description of Event or Problem · 1

A VENTILATOR WAS RETURNED TO THE MFR FOR SERVICE. THERE WAS NO HARM OR INJURY REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
281719 GARBIN 1 VENTILATOR, CONTINUOUS, FACILITY USE CBK RESPIRONICS INC 1058180

Patients

Seq Age Sex Outcome Treatment
1