FDA Adverse Event Malfunction Summary report: N

CASTVAC W/8 FOOT HOSE AND MOBILE STAND

MDR report key: 3192164 · Received June 27, 2013

Report

Report Number
0009616696-2013-00004
Event Type
Malfunction
Date Received
June 27, 2013
Date of Event
June 3, 2013
Report Date
June 5, 2013
Manufacturer
STRYKER INSTRUMENTS-KALAMAZOO
Product Code
LGH
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE REPORTED EVENT, USED FOR WATER AND SPARKS CAME OUT, WAS CONFIRMED. DURING TESTING THE SERVICE TECH OBSERVED THAT THE DEVICE WAS USED FOR WATER, AS VISUAL INSPECTION FOUND THE PC BOARD, MOTOR, ROCKER SWITCH, MID AND BOTTOM CANISTERS, AND RECEPTACLE WERE ALL DAMAGED DUE TO WATER. THE DEVICE IS DESIGNED TO VACUUM DRY MATERIAL ONLY. THE IFU CLEARLY STATES TO NOT USE THE CASTVAC TO VACUUM WET MATERIAL, AND IF USED AS A WET VACUUM THE CASTVAC COULD PRESENT AN ELECTRICAL SHOCK HAZARD. BASED ON THE RISK ASSESSMENT, IFU, AND CASES WITH SIMILAR REPORTED EVENTS THIS FAILURE CAN CAUSE SPARKING AND A POSSIBLE SHOCK TO A USER.

Additional Manufacturer Narrative · 1

THE DEVICE HAS BEEN RECEIVED, BUT THE EVALUATION HAS NOT YET BEGUN. ADDITIONAL INFORMATION MAY BE SUBMITTED ONCE THE QUALITY INVESTIGATION IS COMPLETE.

Description of Event or Problem · 1

IT WAS REPORTED THAT SPARKS CAME OUT OF THE CASTVAC W/8 FOOT HOSE AND MOBILE STAND WHEN IT WAS USED TO SUCTION WATER. THERE WAS NO PATIENT INVOLVEMENT AND NO ADVERSE CONSEQUENCES ASSOCIATED WITH THE DEVICE.

Description of Event or Problem · 1

IT WAS REPORTED THAT SPARKS CAME OUT OF THE CASTVAC W/8 FOOT HOSE AND MOBILE STAND WHEN IT WAS USED TO SUCTION WATER. THERE WAS NO PATIENT INVOLVEMENT AND NO ADVERSE CONSEQUENCES ASSOCIATED WITH THE DEVICE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
291909 CASTVAC W/8 FOOT HOSE AND MOBILE STAND INSTRUMENT, CAST REMOVAL, AC-POWERED LGH STRYKER INSTRUMENTS-KALAMAZOO

Patients

Seq Age Sex Outcome Treatment
1