FDA Adverse Event
Malfunction
Summary report: N
PLV-100
MDR report key: 3192160
·
Received June 20, 2013
Report
- Report Number
- 2518422-2013-01207
- Event Type
- Malfunction
- Date Received
- June 20, 2013
- Date of Event
- September 21, 2010
- Report Date
- September 21, 2010
- Manufacturer
- RESPIRONICS, INC.
- Product Code
- CBK
- PMA / PMN Number
- K832467
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
THE VENTILATOR WAS RETURNED TO THE MFR FOR EVAL AND THE CUSTOMER'S COMPLAINT WAS CONFIRMED. THE SERVICE EVAL FOUND THAT THE BATTERY WAS UNABLE TO REACH FULL CHARGE CAPACITY DUE TO NORMAL WEAR ASSOCIATED WITH RECHARGEABLE BATTERIES. THE DEVICE'S INTERNAL BATTERY WAS REPLACED TO ADDRESS THE ISSUE. THIS REPORT IS BEING SUBMITTED AS PART OF A PROGRAM TO RETROSPECTIVELY REVIEW POSSIBLE DEVICE MALFUNCTIONS THAT DID NOT INVOLVE ANY INJURY, TO DETERMINE WHETHER THEY ARE REPORTABLE UNDER 21 CFR 803 AND THE MFR'S UPDATED REPORTING PROCEDURES. THIS PROGRAM IS BEING UNDERTAKEN TO ADDRESS FDA WARNING LETTER 12-PHI-01.
Description of Event or Problem · 1
THE MFR RECEIVED INFO ALLEGING THE INTERNAL BATTERY WAS NOT LASTING AS LONG AS EXPECTED. THE DEVICE WAS NOT IN PT USE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 281127 | PLV-100 | VENTILATOR, CONTINUOUS, FACILITY USE | CBK | RESPIRONICS, INC. | 35006 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |