FDA Adverse Event Malfunction Summary report: N

CARESCAPE MONITOR B650

MDR report key: 3192157 · Received June 20, 2013

Report

Report Number
9610105-2013-00011
Event Type
Malfunction
Date Received
June 20, 2013
Date of Event
May 21, 2013
Report Date
May 21, 2013
Manufacturer
GE HEALTHCARE FINLAND OY
Product Code
MHX
PMA / PMN Number
K102239
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

PT DATA NOT CURRENTLY AVAILABLE. INVESTIGATION WAS ABLE TO REPRODUCE THE REPORTED ISSUE: STEPS TO REPRODUCE: MONITOR IS PUT TO STANDBY. REMOTE DISCHARGE IS DONE ON THE MONITOR WHILE IT IS IN STANDBY. AFTER THESE STEPS, ALARMS ARE NO LONGER BROADCASTED TO THE UNITY NETWORK. THE SITUATION IS ONLY RECOVERED BY REBOOTING THE MONITOR, OR RETURNING TO STANDBY AND ENDING IT BY SOME OTHER WAY THAN REMOTE DISCHARGE. THERE IS NO PROBLEM WHEN USING THE S5 NETWORK. THE ALARM NETWORK INTERFACE DOES NOT SEND ALARMS TO THE NETWORK WHEN IT THINKS THE MONITOR IT IS STANDBY. THE SOFTWARE ERROR WILL BE CORRECTED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THE B650 MONITOR WITH 2.0.1 SOFTWARE DISCHARGED A PT FROM THE MONITOR REMOTELY FROM A CLINICAL INFO CENTER (CIC) CENTRAL STATION WHILE THE MONITOR WAS ON STANDBY STATE. AFTER A NEW PT WAS ADMITTED IN THE SAME B650 MONITOR, THE ALARMS WERE NOT BROADCASTED TO THE CIC. THERE WAS NO REPORTED PT INJURY ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
281126 CARESCAPE MONITOR B650 MULTI-PARAMETER PATIENT MONITOR MHX GE HEALTHCARE FINLAND OY

Patients

Seq Age Sex Outcome Treatment
1