FDA Adverse Event Malfunction Summary report: N

SYSTEM 7 DUAL TRIGGER ROTARY

MDR report key: 3192143 · Received June 27, 2013

Report

Report Number
0001811755-2013-01474
Event Type
Malfunction
Date Received
June 27, 2013
Date of Event
June 6, 2013
Report Date
June 6, 2013
Manufacturer
STRYKER INSTRUMENTS-KALAMAZOO
Product Code
HWE
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

FAILURE ANALYSIS IS ONGOING; ADDITIONAL INFORMATION MAY BE SUBMITTED ONCE THE RESULTS ANALYSIS IS COMPLETE.

Additional Manufacturer Narrative · 1

THE REPORTED EVENT OF RUN-ON WAS CONFIRMED. DURING FUNCTIONAL TESTING IT WAS OBSERVED THAT THE FORWARD TRIGGER STICKS AND THE DEVICE CONTINUED TO RUN WHEN LET GO, BUT THERE WERE NO FUNCTIONAL ISSUES WITH THE TRIGGER ASSEMBLY. THE STAFF MECHANICAL ENGINEER IN FAILURE ANALYSIS, (B)(4), DETERMINED AN E-BOX FAILURE AS THE PRIMARY FAILURE. THE E-BOX CONTROLS THE ELECTRONICS OF THE DEVICE, AND DAMAGE TO THE ELECTRONIC COMPONENTS WILL CAUSE IT TO NOT FUNCTION PROPERLY IF AT ALL. THIS FAILURE MAY CAUSE RUN-ON OR REDUCED PERFORMANCE OR EFFICIENCY. CORROSION WAS ALSO FOUND INSIDE THE DEVICE, WHICH MAY HAVE CONTRIBUTED TO THE EVENTS AS WELL.

Description of Event or Problem · 1

THE SYSTEM 7 DUAL TRIGGER ROTARY WAS RETURNED TO STRYKER INSTRUMENTS FOR SERVICE, AND DURING FUNCTIONAL EVALUATION IT WAS NOTED THAT THE DEVICE CONTINUED TO RUN AFTER LETTING GO OF THE FORWARD TRIGGER. THERE WAS NO PATIENT INVOLVEMENT AND NO ADVERSE CONSEQUENCES ASSOCIATED WITH THE DEVICE.

Description of Event or Problem · 1

THE SYSTEM 7 DUAL TRIGGER ROTARY WAS RETURNED TO STRYKER INSTRUMENTS FOR SERVICE, AND DURING FUNCTIONAL EVALUATION IT WAS NOTED THAT THE DEVICE CONTINUED TO RUN AFTER LETTING GO OF THE FORWARD TRIGGER. THERE WAS NO PATIENT INVOLVEMENT AND NO ADVERSE CONSEQUENCES ASSOCIATED WITH THE DEVICE.