EON MINI
Report
- Report Number
- 1627487-2013-13937
- Event Type
- Injury
- Date Received
- June 20, 2013
- Date of Event
- May 17, 2013
- Report Date
- May 24, 2013
- Manufacturer
- ST. JUDE MEDICAL - NEUROMODULATION
- Product Code
- LGW
- PMA / PMN Number
- P010032
- Removal / Correction Number
- 1627487-12192011-003-R
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
THIS IPG SERIAL NUMBER WAS INCLUDED IN A FIELD ADVISORY. SJM HAS LIMITED INFORMATION RELATED TO THE PATIENT'S MEDICAL HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PATIENT'S HISTORY TO THE EVENT REPORTED. SJM DEFERS TO THE PATIENT'S PHYSICIAN REGARDING MEDICAL HISTORY.
DEVICE 1 OF 3. REFERENCE MFR. REPORT: 1627487-2013-13938 AND 1627487-2013-13939. IT WAS REPORTED THE PATIENT HAD STATED HER STIMULATION HAD CHANGED APPROXIMATELY A WEEK AGO. X-RAYS REVEALED THE TOP HEADER PORT LEAD WAS SLIGHTLY PULLED OUT OF THE IPG. A SJM REPRESENTATIVE WAS ABLE TO REPROGRAM THE PATIENT USING THE EXISTING BOTTOM PORT LEAD. THE PATIENT WAS TO HAVE A CT/MYELOGRAM SCAN AND FOLLOW-UP WITH HER PHYSICIAN. SURGICAL INTERVENTION MAY BE TAKEN AT A LATER DATE TO ADDRESS THE ISSUE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 280696 | EON MINI | SCS IPG | LGW | ST. JUDE MEDICAL - NEUROMODULATION | 3788 | 3244581 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 48 YR | Other | IMPLANT DATE:| SCS ANCHOR, MODEL 1192 |