MINICAP TRANSFER SET
Report
- Report Number
- 1416980-2013-16517
- Event Type
- Malfunction
- Date Received
- June 27, 2013
- Date of Event
- May 30, 2013
- Report Date
- June 4, 2013
- Manufacturer
- BAXTER HEALTHCARE - MOUNTAIN HOME
- Product Code
- KDJ
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- RS
- Reporter Occupation
- PHYSICIAN
Narratives
(B)(4). THIS COMPLAINT IS NOT CONFIRMED AND THE CAUSE WAS NOT IDENTIFIED BECAUSE THE DEVICE WAS NOT RETURNED FOR EVALUATION. IT WAS REPORTED THAT THE CUSTOMER USED AN ALCOHOL CONTAINING DISINFECTANT TO CLEAN THE TRANSFERS. THE CURRENT INSTRUCTIONS FOR THIS TRANSFER SET INSTRUCT THE USER TO NOT APPLY HYDROGEN PEROXIDE, ALCOHOL OR ANTISEPTIC AGENTS CONTAINING ALCOHOL TO THE CONNECTORS. IF ADDITIONAL RELEVANT INFORMATION BECOMES AVAILABLE, A FOLLOW UP MEDWATCH WILL BE SUBMITTED.
(B)(4). (B)(6). THE DEVICE WAS NOT AVAILABLE FOR EVALUATION. A REVIEW OF ALL BATCH RECORD DOCUMENTS WAS PERFORMED FOR POTENTIALLY ASSOCIATED LOT NUMBER H12D16046 WITH NO ISSUES NOTED DURING THE MANUFACTURING PROCESS. THERE WERE NO DEVIATIONS FROM STANDARD PROCEDURE AND NO EXCEPTIONS RELATED TO THE REPORTED CONDITION WERE NOTED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4).
IT WAS REPORTED THAT WHITE ROLLER CLAMP OF A TRANSFER SET BROKE DURING USE ON A PATIENT. THERE WAS A CLICKING SOUND AND THE LOCK MECHANISM OF THE TRANSFER SET STOPPED WORKING DURING MANUAL PERITONEAL DIALYSIS EXCHANGES. AN ANTISEPTIC COMPOSED OF N-PROPANOL AND PROPANOL-2 WAS BEING USED TO CLEAN THE TRANSFER SET. NO PATIENT INJURY OR MEDICAL INTERVENTION WAS INDICATED IN ASSOCIATION WITH THIS REPORT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 292445 | MINICAP TRANSFER SET | SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE | KDJ | BAXTER HEALTHCARE - MOUNTAIN HOME | H12D16046 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |