UNICEL® DXH 800 COULTER® CELLULAR ANALYSIS SYSTEM
Report
- Report Number
- 1061932-2013-01204
- Event Type
- Malfunction
- Date Received
- June 27, 2013
- Date of Event
- May 30, 2013
- Report Date
- May 30, 2013
- Manufacturer
- BECKMAN COULTER
- Product Code
- GKZ
- PMA / PMN Number
- K081930
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
ADDITIONAL INFORMATION: BECKMAN COULTER RECEIVED RAW DATA FILES FOR A TOTAL OF SEVEN (7) WHOLE BLOOD SAMPLES ALONG WITH THE QUALITY CONTROL (QC) DATA. THE 6C AND LATRON CONTROL RUNS SHOWED NO ALGORITHM ISSUES. ANALYSIS OF THE WHOLE BLOOD RAW DATA DID NOT SHOW ANY ALGORITHM ISSUES EITHER. SOME SAMPLES EXHIBITED HIGH MONOCYTE PERCENT VALUES BUT THE ALGORITHM CORRECTLY SEPARATED THE POPULATION. IN TWO OF THE SAMPLES, A LARGE OVERLAP BETWEEN THE MONOCYTE AND NEUTROPHIL POPULATION IS OBSERVED AND MOST LIKELY CAUSED BY THE PRESENCE OF VERY IMMATURE CELLS; THE ALGORITHM SET SUSPECT MESSAGES FOR BOTH SAMPLES.
A BECKMAN COULTER FSE (FIELD SERVICE ENGINEER) WAS DISPATCHED ON (B)(4) 2013 AND DISCOVERED A LOW DIFFERENTIAL CELL COUNT FOR THE RECENT PATIENT RUNS. THE FSE PROCEEDED TO REPLACE AND PRIME THE DIFFERENTIAL PRESERVATIVE PUMP, PERFORM MTM (MULTI TRANSDUCER MODULE) INTEGRATED GAIN AND PHASE ADJUSTMENTS, AND CALIBRATED THE NRBC (NUCLEATED RED BLOOD CELL) WHOLE BLOOD. THE FSE VERIFIED INSTRUMENT OPERATIONS AND RESULTS MET PUBLISHED SPECIFICATIONS. FAILURE MODE OF THE EVENT IS LIKELY ATTRIBUTED TO THE DIFFERENTIAL PRESERVATIVE PUMP. ALL RELATED MEDWATCH REPORTS ASSOCIATED WITH THIS EVENT: 1061932-2013-01204, 1061932-2013-01283.
THE CUSTOMER REPORTED OBTAINING INTERMITTENT ERRONEOUSLY HIGH BASOPHILS AND/OR MONOCYTES INVOLVING THE UNICEL DXH 800 COULTER CELLULAR ANALYSIS SYSTEM. THE CUSTOMER INDICATED THAT THE ERRONEOUS RESULTS WERE NOT REPORTED OUT OF THE LABORATORY AND THERE WAS NO CHANGE OR AFFECT TO PATIENT TREATMENT IN CONNECTION WITH THIS EVENT. THE CUSTOMER PROVIDED ONLY ONE EXAMPLE OF ERRONEOUS PATIENT RESULTS WITH THE ALLEGED ORIGINAL ERRONEOUS PATIENT RESULTS SHOWING HIGH MONOCYTE (MO%) AND LOW NEUTROPHIL (NE%) RESULTS WHEN COMPARED TO RERUN RESULTS FROM AN ALTERNATE DXH 800 INSTRUMENT. THE CUSTOMER CONSIDERS THE RERUN RESULTS TO BE CORRECT. THIS REPORT REFERENCES THE EVENT WHICH OCCURRED ON (B)(6) 2013; PLEASE SEE MEDWATCH REPORT #1061932-2013-01283 FOR THE REPORT OF THE EVENT WHICH OCCURRED ON (B)(6) 2013.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 292830 | UNICEL® DXH 800 COULTER® CELLULAR ANALYSIS SYSTEM | COUNTER, DIFFERENTIAL CELL | GKZ | BECKMAN COULTER | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 72 YR |