FDA Adverse Event Malfunction Summary report: N

COULTER® LH 750 HEMATOLOGY ANALYZER

MDR report key: 3191878 · Received June 27, 2013

Report

Report Number
1061932-2013-01206
Event Type
Malfunction
Date Received
June 27, 2013
Date of Event
May 31, 2013
Report Date
May 31, 2013
Manufacturer
BECKMAN COULTER
Product Code
GKZ
PMA / PMN Number
K011342
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

A BECKMAN COULTER FSE (FIELD SERVICE ENGINEER) WAS DISPATCHED AND DISCOVERED THAT THE AIR LEAK WAS DUE TO A DISCONNECTED TUBING AT A 5-WAY (FLIPPER RETRACT) SOLENOID FOR THE RIGHT LIFT. THE FSE TRIMMED THE END OF THE TUBING AND RESEATED THE TUBING ONTO THE FITTING TO RESOLVE THE AIR LEAK ISSUE. THE FSE INDICATED THAT THE FLUID LEAK WAS ASSOCIATED WITH A GREEN STRIPE TUBING AT THE SHEATH FLOW COIL AT THE UPPER RIGHT PANEL OF THE DILUTER. THE FSE REPLACED THE GREEN STRIPE TUBING TO RESOLVE THE LEAK AND REPAIRS WERE VERIFIED PER ESTABLISHED PROCEDURES. FAILURE MODE OF THE AIR LEAK IS ATTRIBUTED TO A DISCONNECTED TUBING AT THE FLIPPER RETRACT SOLENOID FOR THE RIGHT LIFT. FAILURE MODE OF THE LEAK IS ATTRIBUTED TO A GREEN STRIPED TUBING AT THE SHEATH FLOW COIL. (B)(4).

Description of Event or Problem · 1

THE CUSTOMER REPORTED A CLEAR FLUID LEAK OF APPROXIMATELY 1 ML FROM THE COULTER LH 750 HEMATOLOGY ANALYZER WHILE TROUBLESHOOTING AN AIR LEAK. THE CUSTOMER INDICATED THAT THE LEAK WAS CONTAINED WITHIN THE INSTRUMENT AND THE INSTRUMENT DID NOT GENERATE ERROR MESSAGES FOR THIS EVENT. THE CUSTOMER WAS WEARING PERSONAL PROTECTIVE EQUIPMENT CONSISTING OF GLOVES, A LABORATORY COAT, AND EYE PROTECTION AT THE TIME OF THE EVENT AND NO INJURY OR EXPOSURE WAS REPORTED. NO ERRONEOUS PATIENT RESULTS WERE GENERATED AND THERE WAS NO CHANGE OR AFFECT TO PATIENT TREATMENT IN CONNECTION WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
294773 COULTER® LH 750 HEMATOLOGY ANALYZER COUNTER, DIFFERENTIAL CELL GKZ BECKMAN COULTER NA NA

Patients

Seq Age Sex Outcome Treatment
1