FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 3191827 · Received June 26, 2013

Report

Report Number
3004209178-2013-95134
Event Type
Malfunction
Date Received
June 26, 2013
Date of Event
June 8, 2013
Report Date
June 8, 2013
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CT
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP WAS UNABLE TO PRIME DURING PRIME TEST DUE TO PROTRUDED/LOOSE DRIVE SUPPORT DISK. NO ALARMS NOTED. THE INSULIN PUMP PASSED DISPLACEMENT TEST. UNABLE TO PERFORM THE OCCLUSION TEST AND EXCESSIVE NO DELIVERY TEST DUE TO PRIME/FILL ANOMALY. THE INSULIN PUMP HAD A SCRATCHED DISPLAY WINDOW AND CRACKED CASE AT DISPLAY WINDOW CORNER.

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED HAVING ISSUES DURING PRIME/REWIND. THE BLOOD GLUCOSE READING WAS 299MG/DL. THE CALLER STATED THAT THE INSULIN PUMP ALARMED, AND THE DRIVE SUPPORT CAP WAS STICKING OUT. ADVISED THE CALLER TO DISCONTINUE THE USE OF THE DEVICE AND REVERT TO BACK UP PLAN. NO FURTHER INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
291008 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-722NAB

Patients

Seq Age Sex Outcome Treatment
1 59 YR