FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 3191798 · Received June 26, 2013

Report

Report Number
3004209178-2013-95090
Event Type
Malfunction
Date Received
June 26, 2013
Date of Event
May 27, 2013
Report Date
June 14, 2013
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IS
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP WAS RECEIVED WITH BLANK DISPLAY DUE TO BROKEN SOLDER JOINT OF POWER CONNECTOR ON THE INTERFACE BOARD. UNABLE TO VERIFY THE ALARM DUE TO THE BLANK DISPLAY. HOWEVER, THE DEVICE HAD A LOOSE DRIVE SUPPORT DISK.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CUSTOMER'S INSULIN PUMP ALARMED. NO FURTHER INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
290999 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-722CAS

Patients

Seq Age Sex Outcome Treatment
1