PARADIGM REAL-TIME INSULIN INFUSION PUMP
MDR report key: 3191798
·
Received June 26, 2013
Report
- Report Number
- 3004209178-2013-95090
- Event Type
- Malfunction
- Date Received
- June 26, 2013
- Date of Event
- May 27, 2013
- Report Date
- June 14, 2013
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IS
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP WAS RECEIVED WITH BLANK DISPLAY DUE TO BROKEN SOLDER JOINT OF POWER CONNECTOR ON THE INTERFACE BOARD. UNABLE TO VERIFY THE ALARM DUE TO THE BLANK DISPLAY. HOWEVER, THE DEVICE HAD A LOOSE DRIVE SUPPORT DISK.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE CUSTOMER'S INSULIN PUMP ALARMED. NO FURTHER INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 290999 | PARADIGM REAL-TIME INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-722CAS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |