FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 3191789 · Received June 26, 2013

Report

Report Number
3004209178-2013-95079
Event Type
Malfunction
Date Received
June 26, 2013
Date of Event
April 9, 2013
Report Date
June 11, 2013
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
EZ
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP WAS RECEIVED WITH A FROZEN DISPLAY AS A RESULT OF A FAULTY COMPONENT ON THE MOTHERBOARD. THE KEYPAD FUNCTIONS PROPERLY WITH TESTING ELECTRONIC UNIT. NO DAMAGE FOUND ON THE KEYPAD TRACE. THE INSULIN PUMP INVOLVED IN THIS EVENT IS THE PARADIGM REAL-TIME VEO INSULIN INFUSION PUMP, WHICH IS NOT MARKETED IN THE UNITED STATES. HOWEVER, THE DEVICE IS SIMILAR TO THE PARADIGM REAL-TIME INSULIN INFUSION PUMP, WHICH IS MARKETED IN THE UNITED STATES.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DISPLAY WAS SHOWING THE NUMBER FIVE, AND IT COULD NOT BE CLEARED BECAUSE THE BUTTON WAS UNRESPONSIVE. IT WAS STATED THAT THE BUTTON WAS NOT PUSHED LONGER THAN THREE MINUTES. NO FURTHER INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
290982 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-554WWL

Patients

Seq Age Sex Outcome Treatment
1