FDA Adverse Event
Malfunction
Summary report: N
PARADIGM REAL-TIME INSULIN INFUSION PUMP
MDR report key: 3191781
·
Received June 26, 2013
Report
- Report Number
- 3004209178-2013-95065
- Event Type
- Malfunction
- Date Received
- June 26, 2013
- Date of Event
- February 2, 2013
- Report Date
- June 10, 2013
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP ALARMED MOTOR ERROR DURING THE BASIC OCCLUSION TEST, AND THE DEVICE WAS UNABLE TO PRIME DURING THE PRIME TEST AS A RESULT OF A PROTRUDED/LOOSE DRIVE SUPPORT DISK. THE DEVICE HAD MISSING END CAP STICKER.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE CUSTOMER'S INSULIN PUMP ALARMED DURING THE MANUAL PRIME PROCESS. THE CUSTOMER'S BLOOD GLUCOSE READING WAS 12.4MMOL/L. NO FURTHER INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 290243 | PARADIGM REAL-TIME INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-722CAL |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |