ENDURANT
Report
- Report Number
- 2953200-2013-01206
- Event Type
- Injury
- Date Received
- June 26, 2013
- Date of Event
- May 16, 2013
- Report Date
- May 28, 2013
- Manufacturer
- MEDTRONIC IRELAND
- Product Code
- MIH
- PMA / PMN Number
- P100021
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- PHYSICIAN
Narratives
(B)(4). METHOD: FILM EVALUATION. RESULTS: OCCLUSION. ANATOMY RELATED, SEVERELY TORTUOUS RIGHT EXTERNAL ILIAC ARTERY. CONCLUSION: OCCLUSION. ANATOMY RELATED, SEVERELY TORTUOUS RIGHT EXTERNAL ILIAC ARTERY.
AN ENDURANT STENT GRAFT SYSTEM WAS IMPLANTED IN A PATIENT FOR THE ENDOVASCULAR TREATMENT OF A 50 MM DIAMETER ABDOMINAL AORTIC ANEURYSM. THE PROXIMAL NECK WAS 19 MM IN DIAMETER AND 25 MM IN LENGTH. THE LEFT COMMON ILIAC ARTERY WAS 19 MM IN DIAMETER AND THE RIGHT COMMON ILIAC ARTERY WAS 14 MM IN DIAMETER. THE RIGHT EXTERNAL ILIAC ARTERY MEASURED 9 MM IN DIAMETER AND THE LEFT EXTERNAL ILIAC ARTERY MEASURED 9 MM IN DIAMETER. AT THE INDEX PROCEDURE THE PHYSICIAN COILED THE RIGHT INTERNAL ILIAC ARTERY AND THEN IMPLANTED THE ENDURANT STENT GRAFTS. THE RIGHT LIMB WAS EXTENDED TO THE RIGHT EXTERNAL ILIAC ARTERY. IT WAS REPORTED THAT ON AN UNKNOWN DATE, IT WAS OBSERVED THAT THE RIGHT LIMB WAS OCCLUDED. THE PHYSICIAN PERFORMED A THROMBECTOMY AND IMPLANTED A BARE STENT IN THE RIGHT LIMB. THE PHYSICIAN STATED THAT THE PATIENT HAD TORTUOSITY OF THE VESSEL AND THERE WAS A POSSIBILITY THAT THE RIGHT LIMB BENT DUE TO VESSEL TORTUOSITY. NO ADDITIONAL CLINICAL SEQUELAE WERE REPORTED AND THE PATIENT IS FINE. A REVIEW OF RETURNED FILMS 7 DAYS POST-IMPLANT SHOWED THAT THE BIFURCATE WAS POSITIONED JUST BELOW THE RENAL ARTERIES. THE AAA MAX DIAMETER WAS 9CM. THE IPSILATERAL LIMB AND 10MM DIAMETER EXTENSION WERE PLACED WITHIN THE RIGHT COMMON ILIAC (ONLY APPROXIMATELY 2CM DISTAL SEAL); THE ID OF THE DISTAL RIGHT LIMB WAS 6-9MM. THE CONTRALATERAL LIMB AND 24MM EXTENSION WERE WITHIN THE FULL LENGTH OF THE LEFT COMMON ILIAC ARTERY. JUST DISTAL TO THE STENT GRAFT LIMBS THE EXTERNAL ILIAC ARTERIES WERE SEVERELY TORTUOUS; BILATERALLY. NO THROMBUS OR ANY OTHER STENT GRAFT ISSUES WERE SEEN. FILMS 3 WEEKS POST-IMPLANT SHOW THAT THE IPSILATERAL (RIGHT) STENT GRAFT COMPONENTS WERE OCCLUDED, BEGINNING JUST BELOW THE BIFURCATE FLOW DIVIDER. NO THROMBUS WAS SEEN WITHIN THE CONTRALATERAL STENT GRAFT LIMBS. NO OBVIOUS CHANGE IN STENT GRAFT IN-VIVO CONFIGURATION WERE SEEN AS COMPARED TO THE EARLIER POST-IMPLANT STUDY. NO STENT GRAFT KINKS WERE OBSERVED. THE ID OF THE DISTAL RIGHT LIMB VARIED FROM 6-8MM. THE EXACT CAUSE OF THE OCCLUSION COULD NOT BE DETERMINED. IT IS POSSIBLE THAT THE SEVERELY TORTUOUS RIGHT EXTERNAL ILIAC ARTERY JUST DISTAL TO THE RIGHT LIMB MAY HAVE CONTRIBUTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 290182 | ENDURANT | SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT | MIH | MEDTRONIC IRELAND | V03991757 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00080 YR | Required Intervention |