FDA Adverse Event Malfunction Summary report: N

MEGA SUTURECUT NEEDLE DRIVERINSTRUMENT

MDR report key: 3191596 · Received June 26, 2013

Report

Report Number
2955842-2013-02313
Event Type
Malfunction
Date Received
June 26, 2013
Date of Event
May 21, 2013
Report Date
May 29, 2013
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. ENGINEERING CONFIRMED THE REPORTED COMPLAINT; HOWEVER, FOR CLARIFICATION THE DAMAGED INSTRUMENT COMPONENT WAS A BROKEN GRIP CABLE. THE GRIP CABLE WAS BROKEN AT THE DISTAL IDLERS. THE IDLER PULLEY SPUN FREELY AND WAS NOT DAMAGED. THE CABLE SEGMENT STUCK OUT AT THE INSTRUMENT'S WRIST. OTHER CABLES AT WRIST WERE NOT DAMAGED. ADDITIONAL FINDING WAS AN INDENTATION AT THE EDGE OF THE DISTAL PULLEY. ENGINEERING CONCLUDED IT WAS MOST LIKELY DUE TO MISHANDLING. ALSO VISIBLE WERE SCRATCHES ON THE SURFACE OF THE PULLEY. THE ENDOWRIST INSTRUMENTS INSTRUCTIONS FOR USE (IFU) SPECIFICALLY STATES: GENERAL PRECAUTIONS AND WARNINGS HANDLE INSTRUMENTS WITH CARE. AVOID MECHANICAL SHOCK OR STRESS THAT CAN CAUSE DAMAGE TO THE INSTRUMENTS. DO NOT USE AN INSTRUMENT TO CLEAN DEBRIS FROM ANOTHER INSTRUMENT INTRAOPERATIVELY. THIS MAY RESULT IN DAMAGE TO THE INSTRUMENTS OR OTHER UNINTENDED CONSEQUENCES, SUCH AS DISCONNECTION OF THE INSTRUMENT TIP. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER; THE REPORTED MALFUNCTION IF TO RECUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A DA VINCI S MYOMECTOMY PROCEDURE A MEGA SUTURECUT NEEDLE DRIVER INSTRUMENT TIP BROKE WHILE SUTURING THE VAGINAL CUFF. THERE WERE NO MISSING PIECES OR FALLEN PIECES REPORTED. THE PLANNED SURGICAL PROCEDURE WAS COMPLETED AND NO PATIENT HARM, ADVERSE OUTCOME, OR INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
290677 MEGA SUTURECUT NEEDLE DRIVERINSTRUMENT ENDOSCOPIC INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420309-01 M10120529 454

Patients

Seq Age Sex Outcome Treatment
1 DA VINCI S SYSTEM INSTRUMENTS AND ACCESSORIES