IVT DISPOSABLE
Report
- Report Number
- 1416980-2013-16494
- Event Type
- Malfunction
- Date Received
- June 26, 2013
- Report Date
- June 10, 2013
- Manufacturer
- BAXTER HEALTHCARE - MALTA
- Product Code
- KPE
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- OTHER
Narratives
(B)(4). IF ADDITIONAL RELEVANT INFORMATION IS OBTAINED, THEN A FOLLOW-UP MDR WILL BE SUBMITTED.
(B)(4). A REVIEW OF ALL BATCH RECORD DOCUMENTS WAS PERFORMED WITH NO ISSUES NOTED DURING THE MANUFACTURING PROCESS. THERE WERE NO DEVIATIONS FROM STANDARD PROCEDURE AND NO EXCEPTIONS RELATED TO THE REPORTED CONDITION WERE NOTED. EVALUATION SUMMARY: THE SAMPLE WAS RECEIVED FOR EVALUATION. VISUAL INSPECTION REVEALED THAT WHITE PARTICLES WERE VISIBLE ON THE BAG, HENCE OUTSIDE THE FLUID PATHWAY. THE CAUSE WAS DETERMINED TO BE A MANUFACTURING ISSUE. UPDATES WERE MADE TO THE MANUFACTURING PROCESS AS A CORRECTIVE ACTION. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A FOLLOW-UP MEDWATCH WILL BE SUBMITTED.
IT WAS REPORTED THAT A 250 ML EVA BAG WAS OBSERVED TO HAVE A PARTICLE INSIDE THE PRIMARY BAG. THIS MALFUNCTION WAS IDENTIFIED BEFORE USE. THERE WAS NO PATIENT INVOLVEMENT. ADDITIONAL INFORMATION WAS REQUESTED AND IS NOT AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 291454 | IVT DISPOSABLE | CONTAINER, I.V. | KPE | BAXTER HEALTHCARE - MALTA | 11J27V259 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |