S-ROM*PIN LOCK,5.0MM DIA
Report
- Report Number
- 1818910-2013-19934
- Event Type
- Injury
- Date Received
- June 26, 2013
- Date of Event
- June 24, 2013
- Report Date
- June 24, 2013
- Manufacturer
- 1219655 DEPUY-RAYNHAM
- Product Code
- JDI
- PMA / PMN Number
- PK963206
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- PHYSICIAN
Narratives
THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED.
CONCLUSION AND JUSTIFICATION STATUS: CONCLUSION AND JUSTIFICATION STATUS: THE COMPLAINT SHALL BE CLOSED WITH AN UNDETERMINED CONCLUSION; IT SHALL BE ENTERED ONTO THE COMPLAINTS DATABASE AND MONITORED THROUGH TREND ANALYSIS. A COMPLAINT SEARCH AND A REVIEW OF THE MANUFACTURING RECORDS WERE CARRIED OUT. NO ANOMALIES OR PREVIOUS COMPLAINTS WERE FOUND RELATING TO THE PROVIDED LOT NUMBER. THE RETURNED X-RAYS AND SURGICAL NOTES WERE REVIEWED BY OUR BIOENGINEERS WHO CONCLUDED AS FOLLOWS: ROOT CAUSE: UNDETERMINABLE, WEAR OF THE POLYETHYLENE HAS OCCURRED AND THIS IS NOT UNEXPECTED GIVEN THE LENGTH OF TIME THE DEVICE HAS BEEN IN THE BODY. THIS WAS A REVISION PROCEDURE WHERE THERE IS A HIGHER RISK OF REVISION THAN FOR A PRIMARY. OSTEOLYSIS IS REPORTED. IT IS UNLIKELY THERE IS A MANUFACTURING FAULT AS THE DEVICE LASTED 18 YEARS. IT IS LIKELY IN THIS CASES THAT AN UNKNOWN COMBINATION OF SEVERAL FACTORS OCCURRED ¿ SUCH AS PATIENT FACTORS, SURGICAL PROCEDURE, SURGICAL PROCESS AND IMPLANT DESIGN. SHOULD ADDITIONAL INFORMATION BE RECEIVED THEN THE COMPLAINT SHALL BE INVESTIGATED FURTHER.
PATIENT PRESENTED IN 2013 WITH PAIN. XRAYS SHOWED POLYETHYLENE LINER WEAR. HIP SHOWED SIGNS OF POLYETHYLENE DISEASE (OSTEOLYSIS).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 291246 | S-ROM*PIN LOCK,5.0MM DIA | PIN LOCK | JDI | 1219655 DEPUY-RAYNHAM | SA102489 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 75 YR | Required Intervention |