FDA Adverse Event Malfunction Summary report: N

VITROS 5600 INTEGRATED SYSTEM

MDR report key: 3191321 · Received June 26, 2013

Report

Report Number
1319681-2013-00132
Event Type
Malfunction
Date Received
June 26, 2013
Date of Event
May 28, 2013
Report Date
June 26, 2013
Manufacturer
ORTHO-CLINICAL DIAGNOSTICS
Product Code
MMI
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INVESTIGATION CONFIRMED THAT A NON-REPRODUCIBLE, HIGHER THAN EXPECTED VITROS TROP I ES RESULT WAS OBTAINED FROM A SINGLE PATIENT SAMPLE WHILE USING THE VITROS 5600 SYSTEM. AN OCD FE PERFORMED SERVICE ACTIONS TO THE WELL WASH SUBSYSTEM OF THE INSTRUMENT AND VERIFIED EXPECTED SYSTEM PERFORMANCE FOLLOWING SERVICE. ADDITIONALLY, THE SAMPLE WAS NOT PROCESSED IN ACCORDANCE WITH THE TUBE MANUFACTURER¿S RECOMMENDATIONS. IT IS LIKELY THAT CELLULAR DEBRIS, DUE TO POOR SAMPLE PREPARATION, WAS PRESENT IN THE AFFECTED SAMPLE, ALTHOUGH THIS COULD NOT BE CONFIRMED. THE INVESTIGATION DETERMINED THAT AN INSTRUMENT ISSUE WAS THE MOST LIKELY CAUSE OF THE EVENT. ADDITIONALLY, A PRE-ANALYTICAL SAMPLE PROCESSING ISSUE CANNOT BE RULED OUT AS A CONTRIBUTING FACTOR. THERE IS NO EVIDENCE TO SUGGEST THAT A REAGENT RELATED ISSUE CONTRIBUTED TO THE EVENT.

Description of Event or Problem · 1

THE CUSTOMER OBTAINED A NON-REPRODUCIBLE, HIGHER THAN EXPECTED VITROS TROP I ES RESULT (0.205 VS. AN EXPECTED RESULT = 0.024 NG/ML) FROM A SINGLE PATIENT SAMPLE PROCESSED ON THE VITROS 5600 SYSTEM. THE AFFECTED SAMPLE RESULT WAS REPORTED OUT OF THE LABORATORY. BIASED RESULTS OF THE MAGNITUDE AND DIRECTION OBSERVED MAY LEAD TO INAPPROPRIATE PHYSICIAN ACTION. IT IS ASSUMED THAT THE CUSTOMER REPEATED THE SAMPLE AS THE HEATH CARE PROVIDER QUESTIONED THE AFFECTED RESULT. A CORRECTED REPORT WAS ISSUED. THERE WAS NO REPORT OF PATIENT HARM AS A RESULT OF THIS EVENT. (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
291203 VITROS 5600 INTEGRATED SYSTEM CHEMISTRY ANALYZER MMI ORTHO-CLINICAL DIAGNOSTICS

Patients

Seq Age Sex Outcome Treatment
1