FDA Adverse Event Summary report: N

RELIANCE SYNERGY WASHER

MDR report key: 3191301 · Received June 26, 2013

Report

Report Number
9680353-2013-00076
Date Received
June 26, 2013
Date of Event
May 27, 2013
Report Date
June 26, 2013
Manufacturer
STERIS CANADA CORPORATION
Product Code
MEC
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

A STERIS SERVICE TECHNICIAN ARRIVED ON SITE AND DURING HIS DISCUSSION OF THE EVENT WITH THE CUSTOMER, HE WAS INFORMED THAT THE LEAK BEGAN DURING THE FILLING PHASE OF THE CYCLE. ADDITIONALLY, WHEN THE PUMP STARTED, THE WATER SPRAYED UNDER THE UNIT.THE TECHNICIAN INSPECTED THE UNIT AND FOUND THE DRYING PISTON VALVE WAS BROKEN. THE TECHNICIAN REPLACED THE DRYING PISTON VALVE AND THE ASSOCIATED HOSE, O-RING AND CLAMP. A TEST CYCLE WAS RUN; THE UNIT WAS CONFIRMED OPERATIONAL AND RETURNED TO SERVICE.

Description of Event or Problem · 1

THE USER FACILITY REPORTED THAT THEIR RELIANCE SYNERGY WASHER WAS LEAKING WATER ONTO THE FLOOR. THE WATER WAS REPORTED TO SPREAD AWAY FROM THE FRONT AND BACK OF THE WASHER. NO PROCEDURAL DELAYS OR CANCELLATIONS OCCURRED. NO INJURIES TO HOSPITAL STAFF OR PATIENTS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
290871 RELIANCE SYNERGY WASHER WASHER MEC STERIS CANADA CORPORATION

Patients

Seq Age Sex Outcome Treatment
1