ULTRACISION HARMONIC ACE
Report
- Report Number
- 3005075853-2013-03307
- Event Type
- Malfunction
- Date Received
- June 26, 2013
- Date of Event
- April 23, 2013
- Report Date
- April 29, 2013
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- LFL
- PMA / PMN Number
- K042777
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KS
- Reporter Occupation
- OTHER
Narratives
(B)(4). INFORMATION ASKED FOR BUT UNKNOWN OR NOT PROVIDED DURING INITIAL CONTACT. SHOULD THE INFORMATION BE PROVIDED LATER, A SUPPLEMENTAL MEDWATCH WILL BE SENT. THE DEVICE WAS RETURNED WITH THE DISTAL TIP OF THE BLADE BROKEN OFF AND IT WAS RETURNED WITH THE DEVICE. THE REMAINING BLADE PORTION WAS SCRATCHED. THE DEVICE WAS FUNCTIONALLY TESTED WITH A GENERATOR AND THE SWITCH BUTTON WAS NOTED TO BE FUNCTIONAL. DURING FUNCTIONAL TESTING ON GEN11 AN INSTRUMENT ERROR WAS DISPLAYED. A PROBABLE CAUSE OF THE DEVICE STOP ACTIVATING AND DISPLAY AN INSTRUMENT ERROR IS BLADE DAMAGE. PROBABLE CAUSES OF BLADE DAMAGE, INCLUDING BREAKAGE, ARE EXTERNAL CONTACT DURING PRE-OP OR GENERAL USE, BLADE CONTACT WITH OTHER DEVICES, STAPLES OR CLIPS DURING THE PROCEDURE OR USING ANY MEANS OTHER THAN THE BLADE WRENCH TO ATTACH OR DETACH THE BLADE. ONCE MINOR BLADE DAMAGE HAS OCCURRED, SUBSEQUENT ACTIVATIONS MAY INCREASE DAMAGE SEVERITY AND RESULT IN AN INSTRUMENT ERROR OR BLADE LOCKOUT LATER IN THE PROCEDURE, AND CONTINUED USAGE CAN RESULT IN A BROKEN BLADE.
IT WAS REPORTED THAT DURING A LAPAROSCOPIC HYSTERECTOMY PROCEDURE THE ACTIVATION BUTTON DID NOT FUNCTION PROPERLY. NO PATIENT CONSEQUENCES REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 290868 | ULTRACISION HARMONIC ACE | INSTRUMENT, ULTRASONIC SURGICAL | LFL | ETHICON ENDO-SURGERY, LLC. | UNK | UNKNOWN |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | GENERATOR, HANDPIECE |