FDA Adverse Event Malfunction Summary report: N

PFNA BLADE PERF L90 SST

MDR report key: 3191225 · Received June 26, 2013

Report

Report Number
8030965-2013-03255
Event Type
Malfunction
Date Received
June 26, 2013
Date of Event
January 26, 2012
Report Date
January 31, 2012
Manufacturer
SYNTHES GMBH
Product Code
KTT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
LE
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

SYNTHES IS SUBMITTING THIS REPORT AS A RESULT OF REMEDIATION ACTIVITIES RELATED TO FDA WARNING LETTER DATED FEBRUARY 2012. DEVICE(S) LISTED IN THIS REPORT IS (ARE) USED FOR TREATMENT, NOT DIAGNOSIS. ANY ADDITIONAL INFORMATION RECEIVED REGARDING THIS EVENT AFTER FILING THIS REPORT SHALL BE FILED ON A SUPPLEMENTAL MDR. (B)(6). AGE AT TIME OF EVENT REPORTED INCORRECTLY IN INITIAL MW. PLACEHOLDER.

Additional Manufacturer Narrative · 1

SYNTHES IS SUBMITTING THIS REPORT AS A RESULT OF REMEDIATION ACTIVITIES RELATED TO FDA WARNING LETTER DATED FEBRUARY 2012. DEVICE(S) LISTED IN THIS REPORT IS (ARE) USED FOR TREATMENT, NOT DIAGNOSIS. ANY ADDITIONAL INFORMATION RECEIVED REGARDING THIS EVENT AFTER FILING THIS REPORT SHALL BE FILED ON A SUPPLEMENTAL MDR. DEVICE IS NOT DISTRIBUTED IN THE UNITED STATES, BUT IS SIMILAR TO DEVICE MARKETED IN THE USA. THE DHR WAS REVIEWED AND NO ISSUES THAT WOULD HAVE RESULTED IN THIS COMPLAINT WERE FOUND. THE ADDITIONAL EVALUATION VISUAL INSPECTION REVEALED THAT THE BLADE WAS RETURNED IN A LOCK POSITION. IT WAS FOUND THAT THE INNER SCREW GOT DAMAGED ON THE TOP EDGE OF THE HEAD AS WELL AS THE INNER HEXAGONAL PART IS BADLY DEFORMED. THEREFORE, IT IS NOT POSSIBLE TO REASSEMBLE THEM ANYMORE. WE CAN ONLY SUPPOSE THAT THE WRONG HANDLING, TIGHTENING UP IN THE WRONG DIRECTION HAS CAUSED THE JAMMING UP.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE BLADE WOULD NOT ATTACH TO SCREWDRIVER. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
290834 PFNA BLADE PERF L90 SST KTT SYNTHES GMBH 2762736

Patients

Seq Age Sex Outcome Treatment
1 88 YR