FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 3191222 · Received June 26, 2013

Report

Report Number
2531779-2013-09046
Event Type
Malfunction
Date Received
June 26, 2013
Report Date
June 1, 2013
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

FOLLOW UP #1 SUBMITTED: 09/11/2013. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. DEVICE EVALUATION: ON EXAMINATION, THE KEYPAD WAS INTACT WITHOUT DAMAGE. ON TESTING, ALL THE KEYPAD BUTTONS HAD NORMAL RESPONSE. THE KEYPAD COVER WAS REMOVED FOR INVESTIGATION AND DID NOT REVEAL ANY EVIDENCE OF DAMAGE, DEFECT OR CONTAMINATION. THE AUDIO BOLUS BUTTON WAS NOTED TO BE TORN. A DAMAGED AUDIO BOLUS BUTTON WILL PERMIT CONTAMINATION TO PERMEATE THE BUTTONS WHICH WILL HAVE A NEGATIVE IMPACT ON BUTTON FUNCTION. THIS SITUATION IS NOT LIKELY TO RESULT IN AN ADVERSE EVENT AS THE DAMAGED AUDIO BOLUS BUTTON SHOULD BE CLEARLY VISIBLE AND PROHIBIT THE USE OF THE KEYPAD BUTTONS. USERS MAY EXPECT KEYPAD DAMAGE DURING NORMAL WEAR AND THE OWNER¿S BOOKLET INSTRUCTS THE USER TO CONTACT CUSTOMER SERVICE IF THE USER SUSPECTS THE PUMP MAY BE DAMAGED. THE AUDIO BOLUS BUTTON COVER WAS REMOVED FOR INVESTIGATION AND REVEALED INTERNAL MOISTURE DAMAGE.

Additional Manufacturer Narrative · 1

THE PUMP HAS BEEN RETURNED TO ANIMAS. EVALUATION HAS NOT YET BEEN COMPLETED. WHEN EVALUATION IS COMPLETE A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSION CAN BE MADE AT THIS TIME.

Description of Event or Problem · 1

ON (B)(6) 2013, THE REPORTER CONTACTED ANIMAS, ALLEGING A BUTTON/KEYPAD (TACTILE CHANGES/UNRESPONSIVE) ISSUE. THE REPORTER STATED THAT THE UP ARROW, DOWN ARROW, AND OK KEYPAD BUTTONS WERE UNRESPONSIVE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
290833 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 38 YR